background image

 

 

can  largely  decrease  the  interference  spark  of  the  equipment,  and  greatly  improve  the  stability  and 

reliability of the system.    Since the driver element is TRIAC, it is suitable for controlling 100-240VAC (not 

for DC power) with current up to 80A.    For the current larger than 80A, an intermediate relay is needed.   

Protection  elements  are  series  wound  to  the  output  terminals,  so  it  can  control  continuous   current  up  to 

0.2A with maximum allowed instantaneous current 2A. 

Relay Module:

 The relay modules are widely used in industrial control.    However, they are the only modules with 

life  time  limit  and  volume  limit  and  have  much  electromagnetic  interference.    There  are  four  types  of 

relay  modules:  L1,  L2,  L3  and  L4.    For  control  output, L1  or L4 large  capacity module  is  recommended.   

L4  is  equipped  with  high  performance  relay  with  small  volume  and  large  capacity  but  it  costs  higher.  L2 

module is small in size. Both of its normal open and normal close terminals have varistor spark absorption 

but  the  capacity  is  small.    It  is  suitable  for  alarm  output.  L1  and  L   are  large  in  size  and  large  capacity. 

For those panel size with 48mm width (including D6, E, F and etc). large size modules cannot be installed 

in  both  PCB  board.  L3  provides  dual  output  to  support  two  loops  of  alarm,  for  example,  AL1+AL2.  If 

mechanical  switch  is  not  preferred  or  L3 is  not  allowed  due  to  size  limitation  mentioned  above ,  G5  (dual 

SSR DC voltage output) with external SSR can be chosen.   

Calibration:

  The  instrument  applies  the  technology  of  automatic  zero  and  digital  calibration,  and  is  free  of 

maintenance.    If  the  error  exceeds  certain  range,  generally,  cleaning  and  drying  the  inside  of  the 

instrument can fix it.    If not, send the instrument back to the factory to examine and repair.  

Warranty  and  maintenance:

  Free  repair  and  maintenance  will  be  given  in  specified  period  of  time  since  the 

delivery.  In  order  to  get  full  and  correct  repair,  please  state  clear  the  phenomena  and  causes  of  the 

malfunction of the instrument. 

 

1.3  Technical Specification 

 

Input type: (universal input) 

Thermocouple: 

K, S, R, E, J, T, B, N, WRe3-WRe25, WRe5-WRe26 etc

 

Resistance temperature detector: Cu50, Pt100 

Linear voltage: 0

5V, 1

5V, 0

1V, 0

100mV, 0

20mV, 0~500mV etc 

Linear current (I4 module in MIO): 0

20mA, 4

20mA, two-wire transmitter etc. 

Extended input: By keeping all  native input specification, another extra specification can be ordered (gradient table 

may be required). 

 

Measurement range 

K(-50

+1300

),  S(-50

+1700

),  R(-50

+1700

),  T(-200

+350

),  E(0

800

),  J(0

1000

),B(200

1800

), N(0

1300

Cu50(-50

+150

), Pt100(-200

+800

Linear Input: -9990

+32000 units defined by user. 

 

Measurement accuracy : 

0.3 (0.3%FS ± 0.1

 

Resolution :

 0.1

  for K, E, J, N, Cu50, Pt100; 1

  for S, R 

 

Temperature shift :

 

0.015%FS /

    (typical value is 70ppm/

 

Sampling cycle : 

A/D converter sampling 8 times per second

 

 

Response time : 

1s ( when digital filter parameter FILt=1) 

 

Alarm  function:

  Four  different  alarm  type,  high limit,  low  limit,  deviation  high limit  and  deviation  low  limit,  optional 

with alarm blocking at the beginning of power on. 

 

Control mode: 

On-off control mode (Hysteresis adjustable) 

Standard PID with auto tuning 

Содержание AI-519

Страница 1: ...ls of temperature pressure flow level humidity etc CONTENTS 1 SUMMARY 1 1 1 MAIN FEATURES 1 1 2 ORDERING CODE DEFINITION 1 1 3 TECHNICAL SPECIFICATION 4 1 4 REAR TERMINAL LAYOUT AND WIRING 6 2 DISPLAY...

Страница 2: ...nge of temperature ISO9001 and CE certified achieving world class level of quality anti interference ability and safety POINTS TO NOTE Always adjust parameters according to input output type and funct...

Страница 3: ...output or SV PV retransmission L1 Large capacity large size normal open relay output module Capacity 250VAC 2A L2 Small capacity small size normal open normal close relay output module Capacity 250VA...

Страница 4: ...ace or install modules by themselves when needed When replacing a module you should pull the controller out of the housing at first insert a small flat tip screwdriver into the opening between the ori...

Страница 5: ...can be chosen Calibration The instrument applies the technology of automatic zero and digital calibration and is free of maintenance If the error exceeds certain range generally cleaning and drying t...

Страница 6: ...power module Linear current output 0 20mA or 4 20mA Output voltage 11V maximum load resistance 500ohm output precision 0 2 FS Electromagnetic compatibility EMC 4KV 5KHz according to IEC61000 4 4 4KV a...

Страница 7: ...ignal can be converted to 1 5V voltage signal by connecting a 250 ohm resistor and input from terminals 11 and 12 3 S S1 or S4 module can be installed in COMM slot for communication If relay TRIAC no...

Страница 8: ...odule with parameter bAud 1 provides MIO on off input simulation SP1 SP2 The external switch should be connected to terminal 3 and 5 Wiring diagram of D6 size instruments 48 48mm 1 Linear voltage of 5...

Страница 9: ...struments 22 5 x 100mm 1 Input 0 5V 1 5V input from 12 11 500mV below input from 10 11 4 20mA with 250ohm shunt resistor converted to 1 5V input from 12 11 2 Fixed with one channel alarm and communica...

Страница 10: ...N4001 Load Thyristor trigger output SCR X2 5 500A BX BX Capacitor Resistor Absorber Circuit Varistor G1 G2 G1 G2 ZNR ZNR V V 100 240VAC TRIAC 5 500A Capacitor Resistor Absorber Circuit Varistor SCR Po...

Страница 11: ...the thermocouple or RTD circuit is break or input specification sets wrong the upper display window will alternately display orAL and the high limit or the low limit of PV and the instrument will aut...

Страница 12: ...password and then press can access System Parameter Table Setpoint Output Magnitude Switch In the basic display status pressing can switch lower display window between displaying setpoint or output m...

Страница 13: ...Generally the smaller the value of CHYS the higher is the precision of auto tuning But the value of CHYS parameter should be large enough to prevent the instrument from error action around setpoint du...

Страница 14: ...utput PV The instrument works as a temperature retransmitter SOP output SV The instrument works as a set current output onoF APId nPId POP SOP Act Acting method rE Reverse acting Increase in measured...

Страница 15: ...t value or corresponding parameter value may be larger than 9999 please avoid choosing 0 but 0 000 instead The display value will become 00 00 when the value is larger than 9999 0 0 0 0 00 0 000 ScL S...

Страница 16: ...he impulse current of induction load CtI 0 5 second and OPrt 5 second 0 3600 second OEF Work range of OPH When PV OEF the upper limit of OUTP is OPL when PV OEF the upper limit of OUTP is 100 For exam...

Страница 17: ...tion are satisfied the instrument should alarm but there is no problem in the system Contrarily in an refrigerating system the unnecessary high limit or deviation high limit alarm may occur at the beg...

Отзывы: