14
any changes, highlight the
ESC-Exit
box and press enter to close the
Auto Stable submenu.
To disable Auto Stable, set the DO% Change input to 0.0.
When Auto Stable is enabled, an
AS
symbol will display next to the
dissolved oxygen value on the run screen and blink during stabilization.
When the dissolved oxygen value has stabilized based on the Auto
Stable settings, the
AS
will display steadily and the instrument will
beep twice if Audio is turned on.
DO UNITS
Highlight
DO Units
and press enter
to open a submenu that allows
you to select the dissolved oxygen units to be displayed on the Run
screen. Highlight the desired unit(s) and press enter to enable or
disable. An enabled dissolved oxygen unit will have an ‘X’ in the box
next to it. Highlight the
ESC-Exit
box and press enter to save any
changes and to close the DO units submenu.
There are three options for displaying dissolved oxygen:
• mg/L
will show DO readings in milligrams per liter on a scale
from 0 to 50 mg/L.
• ppm
(parts per million) is equivalent to mg/L and will show the
DO reading on a scale from 0 to 50 ppm.
• % will show DO readings in a % saturation from 0 to 500%. This
value will be expressed as %L when DO Local% is enabled.
Both % or %L and mg/L or ppm can be displayed simultaneously on
the screen.
TEMPERATURE UNITS
Highlight
Temperature Units
and press enter to open a submenu
that allows you to change the temperature units displayed on the Run
screen. Highlight the desired unit (Celsius or Fahrenheit) and press
enter to enable. The enabled temperature unit will have an ‘X’ in the
box next to it. Only one unit may be enabled at a time. Highlight the
ESC-Exit
box and press enter to save any changes and to close the
Temperature Units submenu.