Alerts and Alarms
130
10
■
Manual Shut-off of a Hazard Alarm
If you respond to a hazard alarm but the PDM cannot communi-
cate with the Pod to acknowledge or shut off the alarm, you can
shut off the alarm manually.
1. Remove the Pod from your body if you have not already
done so.
2. Press the end of a paper clip or similar item straight down into
the manual alarm shut-off port (Figure 10-2) on the bottom
of the Pod. The port can be found by peeling back the adhe-
sive pad from the bottom of the Pod at the square end. The
alarm shut-off port is to the right of the fill port.
3. Press the paper clip straight down into the manual alarm shut
off port until the alarm stops and can no longer be heard.
Then discard the Pod according to local waste disposal regu-
lations.
4. Hard Reset port is located in the PDM’s battery compartment
(Figure 10-3). The “Hard Reset” allows the user to reset the
PDM (i.e. if the screen locks up and goes into an alarm). The
users PDM settings are not lost.
Pod deactivation required.
Press “Ok” to deactivate.
Hazard
Alarm
PDM
Continuous tone.
This alarm occurs after a PDM error
occurs that requires the user to reset the
date and time. Press
OK
to acknowl-
edge.
Change Pod.
Check blood glucose.
Alarm
Type
Source
Beep Pattern
Action
Manual
alarm shut-
off port
Fill
port
Reset
Hard
Figure 10-2
Figure 10-3
Содержание OmniPod my life