5.5.
Tilt + lift interface
5.5.1.
Replace the tilt actuator
Preparation
1.
Set the interface in the maximum tilted position for easy access.
2.
Prepare the wheelchair for maintenance. See section
5.1.
3.
Disconnect the interface cables from the controller.
Procedure
1.
Make sure that the interface is kept in the
tilted position. For instance by placing a
support.
2.
Disconnect the actuator cable.
3.
WARNING!
Make sure that the tilt
mechanism is supported
safely.
Remove the circlips (B) at the pivot (C) at
the end of the actuator.
4.
Remove the old actuator (A).
5.
Put the new actuator into the tilt
mechanism.
6.
Connect the actuator cable.
7.
Make sure that the actuator cable can follow
the movement of the mechanism but that
the cable does not get jammed by the
mechanism.
8.
Check the functionality of the actuator.
B
A
C
D
5.5.2.
Replace lift actuator
Preparation
1.
Set the interface in the maximum lift position for easy access.
2.
Prepare the wheelchair for maintenance. See section
5.1.
3.
Disconnect the interface cables from the controller.
Procedure
1.
Make sure that the interface is kept in the
upper position. For instance by placing a
support.
2.
Disconnect the actuator cable.
3.
WARNING!
Make sure that the lift
mechanism is supported
safely.
Remove the circlips (B) from the pivots (C
and D).
4.
Remove the old actuator (A).
5.
Put the new actuator into the tilt
mechanism.
6.
Connect the actuator cable.
7.
Make sure that the actuator cable can follow
the movement of the mechanism but that
the cable does not get jammed by the
mechanism.
8.
Check the functionality of the actuator.
B
A
C
D
Maintenance tasks & adjustments
24
Service manual - XP
Содержание XP