6-3
IM 01B08D02-02EN
Specifications
6
Communication Signal Specifications
Item
Programmer
Communication
(Used for YSS1000)
RS-485 Communication
Ethernet
Communication
Electrical
Specifications
Complies with RS-232C
Complies with EIA
RS-485
Complies with
IEEE 802.3
10BASE-T/100BASE-TX
Connection
Dedicated front panel
connector
Rear panel screw
terminals
(Five terminals and the
grounding terminal)
Back panel RJ45
connector
Number
connectable
instruments
1 unit
Maximum of 31 units/port
Up to 4-tier cascade
stack (10BASE-T)
Up to 2-tier stack
(100BASE-TX) (*1)
Number of connection: 2
Maximum number of
transactions: 1 (*
3
)
Applicable cable
Dedicated cable A1053UR
(USB/RS-232C
conversion cable)
Shielded twisted-pair cable
Wire size: 0.5 to 1.25 mm
2
(AWG No. 20 to 16)
10BASE-T/100BASE-TX
cable
Cable length
About 2.7 m
Up to 1200m (1.25 mm
2
)
100 m (*2)
*1: Number of cascade connected hubs
*2: Maximum segment length (length between hub and YS1310)
*3: Per a connection