3-1
IM 01B08D02-02EN
Adjusting Direct Inputs (T
emperature/Resistance/Frequency)
3
3.1 List of Direct Input Specifications and Basic
Operations
In the SC Setting Display, communication with the direct input cards, optional
specifications, is made to set or adjust the input specifications.
Examples of sensor type setting, burnout setting, input zero adjustment, and span
adjustment procedures are described in sections 3.2 to 3.4. Other items are also set in
the same way. The table below shows a list of setting items.
No.
Name
Display
Description
Model-basis Data Display
/A01
/A02
/A03
/A04
/A08
01
MODEL
Model
EM1*C
ET5*C
ER5*C
ES1*C
EP3*A
02
TAG NO.
Tag number
16 alphanumeric characters
03
SELF CHK
Self-check results
GOOD or ERROR
A00
DISPLAY
Display item
--
--
--
--
--
A01
INPUT
Input value
[ ][ ][ ].[ ][ ]mV
[ ][ ][ ][ ].[ ]degC [ ][ ][ ].[ ]degC
[ ][ ][ ][ ][ ].[ ]
OHM
[ ][ ][ ][ ]Hz
A02
OUTPUT
Output value
[ ][ ][ ].[ ]%
A03
STATUS
Status
FF
A04
REV NO.
Revision number
n.000 (n: Revision number)
B00
SET
Setting item
--
--
--
--
--
B01
TAG NO1
Tag number 1
8 alphanumeric characters (8 first-half characters of a tag number)
B02
TAG NO2
Tag number 2
8 alphanumeric characters (8 latter half characters of a tag number)
B03
COMMENT1
Comment 1
8 alphanumeric characters (8 first-half characters of comment)
B04
COMMENT2
Comment 2
8 alphanumeric characters (8 latter half characters of comment)
B05
INP TYPE
ER5 input type
--
--
PT/JPT/
PT100-90/
PT50 (Note 1)
--
--
B06
INP TYPE
ET5 input type
--
B/E/J/K/T/R/S/N
--
--
--
B07
LOW CUT
Low cutoff
--
--
--
--
[ ][ ][ ][ ]Hz
(Note 5)
B08
RESIST
ES1 total resistance
--
--
--
[ ][ ][ ][ ][ ].[ ]
OHM
--
B09
UNIT
Temperature unit
(Note 8)
--
degC/degF/K
degC/degF/K
--
--
B10
ZERO
Zero point
[ ][ ][ ].[ ][ ]mV
[ ][ ][ ][ ].[ ]degC [ ][ ][ ].[ ]degC
[ ][ ][ ][ ][ ].[ ]
OHM
[ ][ ][ ][ ]Hz
(Note 5)
B11
SPAN
Span (Note 2)
[ ][ ][ ].[ ][ ]mV
[ ][ ][ ][ ].[ ]degC [ ][ ][ ].[ ]degC
[ ][ ][ ][ ][ ].[ ]
OHM (Note 4)
[ ][ ][ ][ ]Hz
(Note 5)
B12
BURN OUT
Burnout
OFF/UP/DOWN OFF/UP/DOWN OFF/UP/DOWN OFF/UP/DOWN
--
C00
ADJUST
Adjustment item
--
--
--
--
--
C01
OUT 0%
0% output correction
(Note 7)
±10.00
C02
OUT 100%
100% output
correction (Note 7)
±10.00
C03
WIRING R
Burnout correction (Note 3) EXECUTE/RESET EXECUTE/RESET
--
--
--
C04
ZERO ADJ
Input zero adjustment
(Note 6)
[ ][ ][ ].[ ][ ][ ]mV
RST/INC/DEC
[ ][ ][ ].[ ][ ][ ]mV
RST/INC/DEC
[ ][ ][ ].[ ][ ][ ]
OHM
RST/INC/DEC
--
--
C05
SPAN ADJ
Input span adjustment
(Note 6)
[ ][ ][ ].[ ][ ][ ]mV
RST/INC/DEC
[ ][ ][ ].[ ][ ][ ]mV
RST/INC/DEC
[ ][ ][ ].[ ][ ][ ]
OHM
RST/INC/DEC
--
--
C06
ZERO ADJ
Input zero adjustment
(Note 6)
--
--
--
[ ][ ][ ].[ ][ ][ ]
OHM
--
C07
SPAN ADJ
Input span adjustment
(Note 6)
--
--
--
[ ][ ][ ].[ ][ ][ ]
OHM
--
Chapter 3 Adjusting Direct Inputs (Temperature/Resistance/Frequency)
Содержание YS1000 Series
Страница 5: ...Blank Page...
Страница 7: ...vi IM 01B08D02 02EN Chapter 6 Specifications 6 1 General Specifications 6 1 Index Revision Information...
Страница 43: ...Blank Page...
Страница 55: ...Blank Page...
Страница 57: ...Blank Page...
Страница 59: ...Blank Page...