5-6
IM WT310E-02EN
Rated Integration Time
Select the rated integration time.
(Display B)
1.
2.
3
.
Hours
(Display B)
5
.
Press and to set the number.
Press to move the digit.
4.
Minutes
(Display C)
7
.
Seconds
(Display D)
9
.
Confirm the
setting.
Close menu.
The same as
operation 4.
6.
The same as
operation 4.
8.
10.
Note
If you operate this instrument frequently to configure settings such as the D/A output items and rated
integration time, you can connect this instrument to a PC and use WTViewerFreePlus to easily configure
settings from the PC.
Explanation
D/A Output
You can output voltage, current, active power, apparent power, reactive power, power factor, phase
angle, frequency, voltage peak, current peak, and integrated values using a ±5 V FS DC voltage.
Number of Channels
The number of channels varies depending on the specific option that is installed.
• /DA4 option on the WT310E/WT310EH: 4 channels
• /DA12 option on the WT332E/WT333E: 12 channels
Output Format
You can select a preconfigured output format or configure your own original format.
Using the Default Settings
Select one of the default settings to output preconfigured (default) items.
Default Values for Normal Measurement: dFLt-n
Select this setting to output normal measured values. The output settings are as follows:
Suffix Code
/DA4
/DA12
Product Name
WT310E
WT310EH
WT332E
WT333E
Output Channel
ch1
U1
U1
U1
ch2
I1
-
U2
ch3
P1
U3
U3
ch4
fU
UΣ
UΣ
ch5
1
I1
I1
ch6
-
I2
ch7
I3
I3
ch8
IΣ
IΣ
ch9
P1
P1
ch10
-
P2
ch11
P3
P3
ch12
PΣ
PΣ
1 These channels cannot be set.
2 The numbers are used to indicate input elements 1, 2, and 3.
5.3 Producing D/A Output