6-1
IM WT310E-02EN
Troubleshooting, Maintenance, and Inspection
1
2
3
4
5
6
7
App
Index
6.1 Troubleshooting
Faults and Corrective Actions
• If an error code appears on the 7-segment LED display, see section 5.2. Also, for information about
communication error messages, see the appendix of the Communication Interface User’s Manual,
IM WT310E-17EN.
• If servicing is necessary, or if the instrument does not operate properly even after you have
attempted to deal with the problem according to the instructions in this section, contact your nearest
YOKOGAWA dealer.
Problems and Solutions
Refer To
Nothing appears on the 7-segment LED display when you turn on the power.
Securely connect the power cord to this instrument and to the power outlet.
2.3
Set the supply voltage and frequency to within the permitted range.
2.3
The displayed data is not correct.
Confirm that the ambient temperature and humidity are within their specified
ranges.
2.2
Confirm that the display is not being affected by noise.
2.1, 2.5
Check the measurement cable wiring.
2.8 to 2.11
Check the wiring system. (Applies to the WT332E/WT333E.)
2.8 to 2.11,
2.2
1
Confirm that the line filter is off.
2.9
1
Check the measurement period settings.
2.8
1
Check the FAQ at the following URL.
http://tmi.yokogawa.com/
―
Turn the power off and then on again.
2.4
Keys do not work.
Check the REMOTE indicator. If the REMOTE indicator is illuminated, press
SHIFT (LOCAL) to turn it off.
―
Confirm that the keys are not locked.
8.5
1
Unable to make harmonic measurements.
Check the PLL source settings.
6.3
1
Confirm that the input signal that you have selected as the PLL source meets the
specifications.
6.3
1
Unable to configure or control the instrument through the communication interface.
Check to make sure that the GP-IB address setting matches the specifications.
―
2
Check to make sure that the TCP/IP settings match the specifications.
―
2
Check to make sure that the serial (RS-232) interface parameter settings match
the specifications.
―
2
Confirm that the interface meets the electrical and mechanical specifications.
―
2
1 See the User’s Manual, IM WT310E-01EN.
2 See the Communication Interface User’s Manual, IM WT310E-17EN.
Chapter 6
Troubleshooting, Maintenance, and Inspection