6-54
IM WT3001E-17EN
:FLICker:INTerval
Function
Sets the time per measurement of the short-term
flicker value Pst or queries the current setting.
Syntax
:FLICker:INTerval {<NRf>,<NRf>}
:FLICker:INTerval?
<NRf>,<NRf> = 0,30 to 15,00 (measurement
time: minutes, seconds)
Example
:FLICKER:INTERVAL 10,00
:FLICKER:INTERVAL? ->
:FLICKER:INTERVAL 10,00
Description • This setting is available with normal flicker
measurement (
:FLICker:MEASurement
FLICker
). The time per measurement
of dmax caused by manual switching
(
:FLICker:MEASurement DMAX
) is fixed at
1 (min) 00 (sec).
• The setting resolution for the measurement
time is 2 seconds. When an odd number of
seconds is set, it is rounded up to the next
second.
:FLICker:JUDGe
Function
Finishes measurement of dmax caused by
manual switching and performs judgment.
Syntax
:FLICker:JUDGe
Example
:FLICKER:JUDGE
Description This command can be executed with
measurement of dmax caused by manual
switching (
:FLICker:MEASurement DMAX
).
An error occurs if used during normal flicker
measurement (
:FLICker:MEASurement
FLICker
).
:FLICker:MEASurement
Function
Sets the flicker measurement method or queries
the current setting.
Syntax
:FLICker:MEASurement {FLICker|DMAX}
:FLICker:MEASurement?
FLICker = Normal flicker measurement
DMAX = measurement of dmax caused by
manual switching
Example
:FLICKER:MEASUREMENT FLICKER
:FLICKER:MEASUREMENT? ->
:FLICKER:MEASUREMENT FLICKER
:FLICker:EDITion
Function
Sets the IEC standard edition for flicker
measurement or queries the current setting.
Syntax
:FLICker:EDITion {<Edition>}
:FLICker:EDITion?
<Edition> = {ED2P0|ED1P1}
Example
:FLICKER:EDITION ED2P0
:FLICKER:EDITION? ->
:FLICKER:EDITION ED2P0
Description ED2P0: IEC61000-4-15 Ed2.0
ED1P1: IEC61000-4-15 Ed1.1
:FLICker:ELEMent<x>
Function
Sets the target element flicker measurement or
queries the current setting.
Syntax
:FLICker:ELEMent<x> {<Boolean>}
:FLICker:ELEMent<x>?
<x> = 1 to 4 (element)
Example
:FLICKER:ELEMENT1 ON
:FLICKER:ELEMENT1? ->
:FLICKER:ELEMENT1 1
Description When turned ON (1), that element is targeted for
flicker measurement.
:FLICker:FREQuency
Function
Sets the target frequency for flicker measurement
or queries the current setting.
Syntax
:FLICker:FREQuency {<NRf>}
:FLICker:FREQuency?
<NRf> = 50, 60 (target frequency [Hz])
Example
:FLICKER:FREQUENCY 50
:FLICKER:FREQUENCY? ->
:FLICKER:FREQUENCY 50
:FLICker:INITialize
Function
Initializes flicker measurement.
Syntax
:FLICker:INITialize
Example
:FLICKER:INITIALIZE
6.9 FLICker Group