Communication Commands
:STORe:NUMeric?
Function
Queries all settings related to the storage of
numeric data.
Syntax
:STORe:NUMeric?
Example
:STORE:NUMERIC? -> :STORE:NUMERIC:
NORMAL:ELEMENT1 1;ELEMENT2 0;
ELEMENT3 0;ELEMENT4 0;SIGMA 0;
SIGMB 0;U 1;I 1;P 1;S 1;Q 1;
LAMBDA 1;PHI 1;FU 1;FI 1;UPPEAK 0;
UMPEAK 0;IPPEAK 0;IMPEAK 0;CFU 0;
CFI 0;PC 0;TIME 0;WH 0;WHP 0;WHM 0;
AH 0;AHP 0;AHM 0;WS 0;WQ 0;ETA1 0;
ETA2 0;ETA3 0;ETA4 0;F1 0;F2 0;
F3 0;F4 0;F5 0;F6 0;F7 0;F8 0;F9 0;
F10 0;F11 0;F12 0;F13 0;F14 0;
F15 0;F16 0;F17 0;F18 0;F19 0;F20 0
:STORe:NUMeric:NORMal?
Function
Queries all settings related to the stored items of
numeric data.
Syntax
:STORe:NUMeric:NORMal?
Example
:STORE:NUMERIC:NORMAL? ->
:STORE:NUMERIC:NORMAL:ELEMENT1 1;
ELEMENT2 0;ELEMENT3 0;ELEMENT4 0;
SIGMA 0;SIGMB 0;U 1;I 1;P 1;S 1;
Q 1;LAMBDA 1;PHI 1;FU 1;FI 1;
UPPEAK 0;UMPEAK 0;IPPEAK 0;
IMPEAK 0;CFU 0;CFI 0;PC 0;TIME 0;
WH 0;WHP 0;WHM 0;AH 0;AHP 0;AHM 0;
WS 0;WQ 0;ETA1 0;ETA2 0;ETA3 0;
ETA4 0;F1 0;F2 0;F3 0;F4 0;F5 0;
F6 0;F7 0;F8 0;F9 0;F10 0;F11 0;
F12 0;F13 0;F14 0;F15 0;F16 0;
F17 0;F18 0;F19 0;F20 0
:STORe:NUMeric[:NORMal]:ALL
Function
Collectively turns ON/OFF the output of all
element functions when storing the numerical
data.
Syntax
:STORe:NUMeric[:NORMal]:
ALL {<Boolean>}
Example
:STORE:NUMERIC[:NORMAL]:ALL ON
:STORe:NUMeric[:NORMal]:{ELEMent<x>|
SIGMA|SIGMB}
Function
Turns ON/OFF the output of {each
element|ΣA|ΣB} when storing the numeric data.
Syntax
:STORe:NUMeric[:NORMal]:
{ELEMent<x>|SIGMA|SIGMB} {<Boolean>}
:STORe:NUMeric[:NORMal]:
{ELEMent<x>|SIGMA|SIGMB}?
<x> = 1 to 4
Example
:STORE:NUMERIC:NORMAL:ELEMENT1 ON
:STORE:NUMERIC:NORMAL:ELEMENT1? ->
:STORE:NUMERIC:NORMAL:ELEMENT1 1
Description •
:STORe:NUMeric[:NORMal]:SIGMA is
valid
on models with two or more elements.
To turn the output ON, wiring unit ΣA must exist
by setting the wiring system beforehand using
the
[:INPut]WIRing
command.
•
:STORe:NUMeric[:NORMal]:SIGMB
is
valid on models with four elements. To turn
the output ON, wiring unit ΣB must exist by
setting the wiring system beforehand using the
[:INPut]WIRing
command.
:STORe:NUMeric[:NORMal]:PRESet<x>
Function
Presets the output ON/OFF pattern of the
element function for storing the numeric data.
Syntax
:STORe:NUMeric[:NORMal]:PRESet<x>
<x> = 1 to 2 (preset pattern number)
Example
:STORE:NUMERIC:NORMAL:PRESET1
Description For details on the storage pattern when preset is
executed, see the
User’s Manual IM WT3001E-
01EN
.
:STORe:NUMeric[:NORMal]:<Function>
Function
Turns ON/OFF the output of the function when
storing the numerical data or queries the current
setting.
Syntax
:STORe:NUMeric[:NORMal]:<Function>
{<Boolean>}
:STORe:NUMeric[:NORMal]:<Function>?
<Function> =
{U|I|P|S|Q|...}
(See the
function selection list (1) of “DISPlay group” on
page 6-44.)
Example
:STORE:NUMERIC:NORMAL:U ON
:STORE:NUMERIC:NORMAL:U? ->
:STORE:NUMERIC:NORMAL:U 1
6.21 STORe Group