6-71
IM WT3001E-17EN
1
2
3
4
5
6
7
App
Index
Communication Commands
[:INPut]:SCALing[:STATe]:ELEMent<x>
Function
Turns ON/OFF the scaling of the element or
queries the current setting.
Syntax
[:INPut]:SCALing[:STATe]:
ELEMent<x> {<Boolean>}
[:INPut]:SCALing[:STATe]:
ELEMent<x>?
<x> = 1 to 4 (element)
Example
:INPUT:SCALING:STATE:ELEMENT1 OFF
:INPUT:SCALING:STATE:ELEMENT1? ->
:INPUT:SCALING:STATE:ELEMENT1 0
[:INPut]:SYNChronize?
Function
Queries the synchronization source of all
elements.
Syntax
[:INPut]:SYNChronize?
Example
INPUT:SYNCHRONIZE? ->
:INPUT:SYNCHRONIZE:ELEMENT1 I1;
ELEMENT2 I2;ELEMENT3 I3;ELEMENT4 I4
[:INPut]:SYNChronize[:ALL]
Function
Collectively sets the synchronization source of all
elements.
Syntax
[:INPut]:SYNChronize[:ALL] {U<x>|
I<x>|EXTernal|NONE}
<x> = 1 to 4 (element)
EXTernal = External clock input (Ext Clk)
NONE = No synchronization source
Example
:INPUT:SYNCHRONIZE:ALL I1
[:INPut]:SYNChronize:ELEMent<x>
Function
Sets the synchronization source of the element or
queries the current setting.
Syntax
[:INPut]:SYNChronize:
ELEMent<x> {U<x>|I<x>|EXTernal|
NONE}
[:INPut]:SYNChronize:ELEMent<x>?
<x> = 1 to 4 (element)
EXTernal = External clock input (Ext Clk)
NONE = No synchronization source
Example
:INPUT:SYNCHRONIZE:ELEMENT1 I1
:INPUT:SYNCHRONIZE:ELEMENT1? ->
:INPUT:SYNCHRONIZE:ELEMENT1 I1
[:INPut]:SYNChronize:{SIGMA|SIGMB}
Function
Collectively sets the synchronization source of all
elements belonging to wiring unit {ΣA|ΣB}.
Syntax
[:INPut]:SYNChronize:{SIGMA|
SIGMB} {U<x>|I<x>|EXTernal|NONE}
Example
:INPUT:SYNCHRONIZE:SIGMA I1
Description •
[:INPut]:SYNChronize:SIGMA
is valid
only on models with 2 to 4 elements.
•
[:INPut]:SYNChronize:SIGMB
is valid
only on models with 4 elements.
• This command is invalid, if the wiring unit
{ΣA|ΣB} does not exist as a result of the wiring
system setting (
[:INPut]:WIRing
).
[:INPut]:VOLTage?
Function
Queries all settings related to the voltage
measurement.
Syntax
[:INPut]:VOLTage?
Example
:INPUT:VOLTAGE? -> :INPUT:VOLTAGE:
RANGE:ELEMENT1 1.000E+03;
ELEMENT2 1.000E+03;
ELEMENT3 1.000E+03;
ELEMENT4 1.000E+03;:INPUT:VOLTAGE:
AUTO:ELEMENT1 0;ELEMENT2 0;
ELEMENT3 0;ELEMENT4 0;:INPUT:
VOLTAGE:MODE:ELEMENT1 RMS;
ELEMENT2 RMS;ELEMENT3 RMS;
ELEMENT4 RMS
[:INPut]:VOLTage:AUTO?
Function
Queries the voltage auto range setting (ON/OFF)
of all elements.
Syntax
[:INPut]:VOLTage:AUTO?
Example
:INPUT:VOLTAGE:AUTO? ->
:INPUT:VOLTAGE:AUTO:ELEMENT1 0;
ELEMENT2 0;ELEMENT3 0;ELEMENT4 0
[:INPut]:VOLTage:AUTO[:ALL]
Function
Collectively turns ON/OFF the voltage auto range
of all elements.
Syntax
[:INPut]:VOLTage:AUTO
[:ALL] {<Boolean>}
Example
:INPUT:VOLTAGE:AUTO:ALL ON
[:INPut]:VOLTage:AUTO:ELEMent<x>
Function
Turns ON/OFF the voltage auto range of the
element or queries the current setting.
Syntax
[:INPut]:VOLTage:AUTO:
ELEMent<x> {<Boolean>}
[:INPut]:VOLTage:AUTO:ELEMent<x>?
<x> = 1 to 4 (element)
Example
:INPUT:VOLTAGE:AUTO:ELEMENT1 ON
:INPUT:VOLTAGE:AUTO:ELEMENT1? ->
:INPUT:VOLTAGE:AUTO:ELEMENT1 1
[:INPut]:VOLTage:AUTO:{SIGMA|SIGMB}
Function
Collectively turns ON/OFF the voltage auto range
of all elements belonging to wiring unit {ΣA|ΣB}.
Syntax
[:INPut]:VOLTage:AUTO:{SIGMA|SIGMB}
{<Boolean>}
Example
:INPUT:VOLTAGE:AUTO:SIGMA ON
Description •
[:INPut]:VOLTage:AUTO:SIGMA
is valid
only on models with 2 to 4 elements.
•
[:INPut]:VOLTage:AUTO:SIGMB
is valid
only on models with 4 elements.
• This command is invalid, if the wiring unit
{ΣA|ΣB} does not exist as a result of the wiring
system setting (
[:INPut]:WIRing
).
6.14 INPut Group