IM 12D06A03-01E
Appendix 3-7
2/4
QIS 12D06A03-21E
x
When the message display shows “4 mA,” the output current must be within in the range
shown in Table 1.
x
Press the [ENT] key repeatedly until the message displays shows “12 mA.” When it shows
“12 mA,” the output current must be within the range shown in Table 1.
To skip the current output not needed to be checked, just press the [ENT] key.
x
Press the [ENT] key repeatedly until the message displays shows “20 mA.” When it shows
“20 mA,” the output current must be within the range shown in Table 1.
x
From the above last step, proceed directly to the temperature indication check in Item 3.3.
Table 1
Current Output (Indication)
Output Range
4.0 mA
4 ±0.02 mA DC
12.0 mA
12 ±0.02 mA DC
20.0 mA
20 ±0.02 mA DC
3.4 Temperature
Indication
Check
<Test>
x
From the last step in Item 3.3, Current Output Test, press the [ENT] key repeatedly until the
message display shows “PT1000.”
x
Change the resistance value of the resistance box 1 (for temperature) as shown in Table 2-1
and check the temperature readings. Each temperature reading must be within the range.
Table 2-1 (PT1000)
Resistance of
Resistance Box 1
Indication range
1097.3
:
25.0
±0.3°C
1385.0
:
100.0
±0.3°C
x
Press the [ENT] key. The message display will show “30k NTC.”
x
Change the resistance value of the resistance box 1 (for temperature) as shown in Table 2-2
and check the temperature readings. Each temperature reading must be within the range.
Table 2-2 (30k NTC)
Resistance of
Resistance Box 1
Indication range
30
k
:
25.0
±0.3°C
2.069
k
:
100.0
±0.3°C
x
From the above last step, proceed directly to the resistance (conductivity) indication check in
Item 3.5.
3.5 Resistance (Conductivity) Indication Check
<Test>
x
From the last step in Item 3.4, Temperature Indication Check, press the [ENT] key
repeatedly until the message display shows “SEL.10.”
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.1.”
x
Set resistance box 2 to the value shown in Table 3 and check the indication. The resistance
indication must be within the range.
Содержание vigilantplant EXA ISC202G
Страница 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 12: ...IM 12D06A03 01E 1 4 Introduction...
Страница 53: ...IM 12D06A03 01E Parameter setting 5 11 5 2 4 Temperature compensation for first conductivity value To TEMP 2...
Страница 54: ...IM 12D06A03 01E 5 12 Parameter setting 5 2 5 Temperature Compensation for second conductivity value To SERV...
Страница 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D06A03 01E 2 20 Appendix...
Страница 97: ...IM 12D06A03 01E Appendix 3 5...
Страница 102: ...IM 12D06A03 01E 3 10 Appendix...
Страница 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Страница 106: ...IM 12D06A03 01E 3 14 Appendix...
Страница 110: ...IM 12D06A03 01E 3 18 Appendix...