IM 12D06A03-01E
Appendix 2-1
12. APPENDIX 2
12-1. Preface
This appendix contains these items.
1. Method of wiring and parameter setting the following:.
PH201G*B Dedicated Distributor
BA2O
Junction Terminal Box
WF1OJ Extension
Cable
2. Quick reference for parameter setting
3. Installation factor adjustment.
To ensure that this measurement system can be operated safely and also exhibit its full performance, be
sure to read this appendix before use.
This appendix does not describe SDBT Distributor which is the component unit of the ISC2O2 two-wire
inductive conductivity transmitter system. This unit comes with an instruction manual, so read the instruc-
tion manual IM 1B04T02-01E for details of the unit concerned.
12-2. Wiring diagrams
14
13
Output
(1-5V DC)
A(+)
B(-)
(+)C
(-)D
17
+
-
G
11
12
16
15
Inductive Conductivity
transmitter ISC202G
When PH201G (Style B)
distributor is used
*1
Output
(1-5V DC)
1(+)
2(-)
(+)A
(-)B
SDBT
distributor
*1
Class D (100 ohm or less)
ground
*
2
b
a
d
c
HOLD FAIL
Output
(1-5V DC)
(+)F
(-)H
Relay contact
Non-explosionproof system
Inductive Conductivity
detector ISC40GJ
Output
(1-5V DC)
(+)F
(-)H
*1 Use two-wire cable with OD (Outside Diameter) of 6 to 12 mm
Make sure distributor voltage does not drop below minimum for ISC202G
*2 Ground ISC202G ( Class D ground: 100 ohm or less)
Junction Terminal
Box BA20
Extension Cable
WF10J
Содержание vigilantplant EXA ISC202G
Страница 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 12: ...IM 12D06A03 01E 1 4 Introduction...
Страница 53: ...IM 12D06A03 01E Parameter setting 5 11 5 2 4 Temperature compensation for first conductivity value To TEMP 2...
Страница 54: ...IM 12D06A03 01E 5 12 Parameter setting 5 2 5 Temperature Compensation for second conductivity value To SERV...
Страница 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D06A03 01E 2 20 Appendix...
Страница 97: ...IM 12D06A03 01E Appendix 3 5...
Страница 102: ...IM 12D06A03 01E 3 10 Appendix...
Страница 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Страница 106: ...IM 12D06A03 01E 3 14 Appendix...
Страница 110: ...IM 12D06A03 01E 3 18 Appendix...