IM 12D06A03-01E
3-8 Appendix
3/4
QIS 12D06A03-21E
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.2.” There is no need to check here, so just press the [ENT] key.
x
The message display shows “
WAIT
” momentarily, followed by “RES.3.”
x
Set resistance box 2 to the value shown in Table 3 and check the indication. The resistance
indication must be within the range.
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.4.” There is no need to check here, so just press the [ENT] key.
x
The message display shows “
WAIT
” momentarily, followed by “RES.5.”
Table 3 SEL.10
Indication
Setpoint of Resistance Box 2
Indication Range
RES. 1
30.00
:
0.300±0.002
RES. 3
800.00
:
8.00±0.04
RES. 5
30.00k
:
300±2
x
Change the number of turns of wire onto the ISC40GJ sensor from ten to one.
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.6.”
x
Set resistance box 2 to the value shown in Table 4 and check the indication. The resistance
indication must be within the range.
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.7.” There is no need to check here, so just press the [ENT] key.
x
The message display shows “
WAIT
” momentarily, followed by “RES.8.”
x
Set resistance box 2 to the value shown in Table 4 and check the indication. The resistance
indication must be within the range.
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.9.”
x
Set resistance box 2 to the value shown in Table 4 and check the indication. The resistance
indication must be within the range.
x
After the test, press the [ENT] key. The message display shows “READY.”
x
Press the [ENT] key to restart the transmitter.
Table 4 SEL1
Indication
Setpoint of Resistance Box 2
Indication Range
RES. 6
300.0
:
300±2
RES. 8
8.000k
:
8.00±0.10k
RES. 9
80.00k
:
80.0±6.3k
Содержание vigilantplant EXA ISC202G
Страница 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 12: ...IM 12D06A03 01E 1 4 Introduction...
Страница 53: ...IM 12D06A03 01E Parameter setting 5 11 5 2 4 Temperature compensation for first conductivity value To TEMP 2...
Страница 54: ...IM 12D06A03 01E 5 12 Parameter setting 5 2 5 Temperature Compensation for second conductivity value To SERV...
Страница 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D06A03 01E 2 20 Appendix...
Страница 97: ...IM 12D06A03 01E Appendix 3 5...
Страница 102: ...IM 12D06A03 01E 3 10 Appendix...
Страница 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Страница 106: ...IM 12D06A03 01E 3 14 Appendix...
Страница 110: ...IM 12D06A03 01E 3 18 Appendix...