IM 12D06A03-01E
Appendix 1-3
Primary choices
Defaults
Alternatives
Reference Menu or
on page
Service code (SC)
Measurement Conductivity
1st Temp. compensation
NaCl in water
Fixed T.C., Matrix
5-2-4
‘Commissioning’
2nd Temp. compensation
NaCl in water
Fixed T.C., Matrix (inactive)
5-2-5
‘Commissioning’
2nd Line display
Process temp.
1st compensation method, Weight%,
5-1-3/4
‘Maintenance’
Output (mA), C.C., Ref. Temp., Software
release, 2nd compensated Conductivity,
2nd compensation method
Range (Linear)
0-1000 mS/cm
0-1999 mS/cm
5-2-1
‘Commissioning’
Temperature compensator
30k NTC
Pt 1000
5-3-2
SC 10
Temperature unit Sensor
Celsius (ºC)
Fahrenheit ( ºF )
5-3-2
SC 11
Cell constant
1.88 /cm
Any value between 0.2 and 19.99 /cm
5-3-1
SC 03
Communication
HART® enabled
disable HART®, PH201*B
5-7
SC 60
Burn out
inactive
HI or LO, Pulse burnout
5-5
SC 32
HOLD during maintenance inactive
Hold last value or fixed value
5-2-2
‘Commissioning’
Calibration temperature
inactive
Adju/- 15ºC
5-3-2
SC 12
AIR (zero) calibration
inactive
Adju/- 15
μ
S/cm 5-3-1
SC
04
C.C. Calibration
inactive
0.2 and 19.99 /cm
5-1-2
‘Maintenance’
Diagnostics
Hard alarm (all errors) Hard or soft choices
5-6
SC 53
Instrument Limit: E5 (high)
3S
E5 (High) 0.10 - 9.99 S
5-6
SC 54
Instrument Limit: E6 (low)
0.5
μ
S
E6 (Low) 0.00 - 99.9
μ
S 5.6
SC
54
Password protection
inactive
(In)active password for different levels
5-6
SC 52
Output in Concentration units inactive
Linearization of output, Weight% on LCD 5-6
SC 55
Code Display
Default Values
Setting #1
Setting #2
Setting #3
03 *C.C. 1.88
/cm
10
*T.SENS
0 : 30 kNTC
11
*T.UNIT
0 : ºC
12 *T.ADJ 0
ºC
20 *T.R.
ºC 25ºC
21
*T.C.1
2.1% per ºC
*T.C.2 2.1%
per
ºC
22
*MATRX
1 : H
2
SO
4
31 *OUTP.F 0:
Linear
32
*BURN
0: No Burnout
50 *RET 1:
On
52 *PASS 0.0.0.:
off.off.off
53
*Err. 05
1 : HARD
*Err. 06
1 : HARD
*Err. 07
1 : HARD
*Err. 08
1 : HARD
54 *E5.L 3S
*E6.LIM
5
μ
S
*OUTP
0% : 0
μ
S
100% : 1000mS
*HOLD
disabled
*TEMP.1
NaCl
*TEMP.2
NaCl
60
*COMM.
1.0: On, write enable
mA
mA
mA
mA
mA
mA
mA
mA
11-5. Coded service settings (default)
11-4. Configuration Checklist For ISC202
Содержание vigilantplant EXA ISC202G
Страница 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 12: ...IM 12D06A03 01E 1 4 Introduction...
Страница 53: ...IM 12D06A03 01E Parameter setting 5 11 5 2 4 Temperature compensation for first conductivity value To TEMP 2...
Страница 54: ...IM 12D06A03 01E 5 12 Parameter setting 5 2 5 Temperature Compensation for second conductivity value To SERV...
Страница 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D06A03 01E 2 20 Appendix...
Страница 97: ...IM 12D06A03 01E Appendix 3 5...
Страница 102: ...IM 12D06A03 01E 3 10 Appendix...
Страница 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Страница 106: ...IM 12D06A03 01E 3 14 Appendix...
Страница 110: ...IM 12D06A03 01E 3 18 Appendix...