IM 12D06A03-01E
Parameter setting 5-17
Code Display
Function
Function
detail
X
Y
Z
Default
values
mA Outputs
30
Not
used
31
*OUTP.F
mA output functions
Linear
0
0 Linear
Table
1
32
*BURN
Burn function
No burnout
0
0 No Burn.
Burnout
downscale
1
Burnout
upscale
2
Pulse
burnout
3
33, 34
Not used
35
*TABLE
Output table for mA
*0%
Linearisation table for mA in 5% steps.
*5%
The measured value is set in the main
*10%
display using the
keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value may
*95%
be skipped, and a linear interpolation will
*100%
take
place.
36-49
Not
used
mA
Concentration Output function is done in de following order:
• Set *OUTP.F. (Service Code 31) to table
• Set the Concentration range in % (Service Code 55)
• Set table values (%output and Conductivity values) in *TABLE (Service Code 35)
Содержание vigilantplant EXA ISC202G
Страница 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 12: ...IM 12D06A03 01E 1 4 Introduction...
Страница 53: ...IM 12D06A03 01E Parameter setting 5 11 5 2 4 Temperature compensation for first conductivity value To TEMP 2...
Страница 54: ...IM 12D06A03 01E 5 12 Parameter setting 5 2 5 Temperature Compensation for second conductivity value To SERV...
Страница 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Страница 92: ...IM 12D06A03 01E 2 20 Appendix...
Страница 97: ...IM 12D06A03 01E Appendix 3 5...
Страница 102: ...IM 12D06A03 01E 3 10 Appendix...
Страница 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Страница 106: ...IM 12D06A03 01E 3 14 Appendix...
Страница 110: ...IM 12D06A03 01E 3 18 Appendix...