Standard Specifications 

IM 01G05B03-01E 1st Edition 

 

Model and Suffix Code 

Ultrasonic flowmeter 

Model Suffix 

code 

Specification

US300FM ····························· 

Ultrasonic 

flowmeter 

Output 

-A1 ······················ 

One current output 

 

-A2 ······················ 

Two current outputs 

Power Supply 

1 ··················· 

100 to 230 V AC ±10% 

 

4 ··················· 

24 V DC ±10% 

Input Channel 

-1 ················ 

One input channel   
(one-path) 

-2 ················ 

Two input channels   
(dual-path) 

Electrical connection 

-4 ··········· 

ISO M20 x 1.5, female 

Option 

/PU1 ···· 

One binary (pulse or alarm) 
output (open-collector) (*1)

 

/PU2 ···· 

Two binary (pulse or alarm) 
outputs (open-collector) (*1)

 

/FQ1 ···· 

Frequency output (open-
collector, 0 to 1 kHz) (*2) 

 

/BGT ···· 

Tag number on nameplate 
(in the nameplate label, 
maximum 16 characters) 

 

/SCT ···· 

Tag number on stainless 
steel tag plate (maximum 16 
characters) 

 

/PMT ···· 

Pipe mounting fixture 

 
*1: Option /PU1 and /PU2 are exclusive. 
*2: Option /FQ1 is not selectable for two current output (-A2) 

models 

 

Connection cable (*8) 

Model Suffix 

code 

Specification

US300FC ····················

Connection cable 

Length 

-Gxxx ········· 

xxx: Cable length 001 to 300 [m] 

 
*8: Two sets of connection cables are necessary when 

applying dual channel/path measurement. 

 

Data transfer software 

Model Suffix 

code 

Specification

US300SA ····················

Data transfer software (Windows 
version) Including connecting kit 
(RS232 cable for connection, RS232 
adapter 9/25) 

Language 

-1 ················ 

English / German version 

 

00 ······· 

Always 00 

 

Transducers (*3) 

Model

Suffix code

Specification

US300FT

·························· 

Transducers for fixed type 

Usage (*4)

-G ····················· 

General purpose (IP65) 

-W ···················· 

Waterproof (IP67) 

Pipe Size / 
Fluid 
Temperature 
(*5) 

BG ················ 

Medium & General (with 3 m / 
9.8 ft cable) 

BH ················ 

Medium & High (with 3 m / 9.8 ft 
cable) 

CG ················ 

Large & General (with 4.4 m / 
14.4 ft cable) 

CH ················ 

Large & High (with 4.4 m / 14.4 
ft cable) 

DG ················ 

Very large & General (with 12 m / 
39.4 ft cable) 

 

-N ·············· 

Always N 

Fixing band, strap 
and clips (*6) 

B ············ 

For 25 to 1,400 mm (1 to 55 in.)

Two fixing bands   
One strap of 10 m (32 ft) 
length 
Two clips of medium type 
Two clips of large type 

C ············ 

For 1,400 to 2,800 mm (55 to 
110 in.) 

One strap of 20 m (65 ft ) 
length 
Two clips of large type 

D ············ 

For 2,800 to 6,500 mm (110 to 
255 in.) 

Two straps of 20 m (65 ft) 
length 
Two clips of large type 

N ············ 

None 

Acoustic couplant 
(*7) 

G ········· 

General temperature type (non-
adhesive, –30 to +130°C / –22 
to +266°F) 

H ········· 

High temperature type (non-
adhesive, –30 to +200°C / –22 
to +392°F) 

R ········· 

Weatherproof type (adhesive, 
–40 to +180°C / –40 to +356°F)

N ········· 

None 

Option 

/TTP ··· 

Transducer tag plate (maximum 
16 characters) 

 
*3: Two sets of transducers are necessary when applying dual 

channel/path measurement. 

*4: A Tokuchu request is necessary when transducers are 

installed where condensation occurs. However the pipe 
size must be over 100 mm (Large type or Very large type). 

*5: The alphabetic characters in the suffix code represent pipe 

sizes and fluid temperature ranges are below. 
B: Medium type (25 to 400 mm / 1 to 16 in.) 
C: Large type (100 to 2,500 mm / 4 to 98 in.) 
D: Very large type (2,000 to 6,500 mm / 78 to 255 in.) 
G: General temperature (–30 to +130°C / –22 to +266°F) 
H: High temperature (–30 to +200°C / –22 to +392°F) 

*6: When the pipe size is less than 50 mm (2 in.), select 

N(None) here first and order separately one each of 
USPA053 mounting fixture short ruler type, USPA036 
fixing chains (0.5 m/1.6 ft length), and USPA097 couplant 
weatherproof type. It will avoid any stress that could be 
caused by expansion of the pipe after installation. The 
couplant weatherproof type can be specified by the code R 
for couplant.   

Содержание US300FM

Страница 1: ...ite White Perform soldering Finish the brown cables as well as white cables You do not connect the thin lines to any terminals Brown 12 2 Less than 10mm 12 2 Perform soldering Shield line Cable shield...

Страница 2: ...ccurs A Tokuchu request is necessary for Submersible type However the diameter must be over 100 mm Changes to the section 3 5 5 Connection to the Power Supply page 20 AC power supply DC power supply T...

Страница 3: ...y this behavior Addition and deletion to the section 4 1 1 Key Operations page 23 and related pages Refer to the following table for the function of corresponding key operations The correct explanatio...

Страница 4: ...a wall thickness probe or some other ways and avoid mounting the transducers on such parts Attention Instructions for using Acoustic couplant weatherproof type Shin Etsu Silicone 1 COMPONENT RTV are a...

Страница 5: ...the section 14 5 Activation of a Pulse Output page 95 Maximum pulse output rate is 2 pulse per second 20 Set the Pulse Value and the Pulse Width so that the actual maximum flow in the pipe is less th...

Страница 6: ...Measuring Input One or two dual Channel A Channel B Current Output One or two Range 4 to 20 mA Load resistance 0 to 500 Note The current outputs may temporarily turn unstable during the power on seque...

Страница 7: ...tion only for test service maintenance purposes Time Programmable Measurement Function Automatic start stop of the measurement without human operation using internal clock for specific systems Can be...

Страница 8: ...suring data set in graphic format Marker type and color for each line of values selectable Scales for time axis and value axis can be changed from default condition of automatic scaling Graph printing...

Страница 9: ...th 4 4 m 14 4 ft cable CH Large High with 4 4 m 14 4 ft cable DG Very large General with 12 m 39 4 ft cable N Always N Fixing band strap and clips 6 B For 25 to 1 400 mm 1 to 55 in Two fixing bands On...

Страница 10: ...g fixture short ruler type for transducers medium pipe size type temperature range 30 to 200 C 22 to 392 F set of two blocks with 120 mm 4 7 in ruler USPA054 Mounting fixture standard type for transdu...

Страница 11: ...flowmeter US300FM Unit mm inch Transducers US300FT B Transducers US300FT C US300FT D Transducers X m inch Y m inch X Y m inch Weight kg lb US300FT B 2 0 78 7 1 0 39 4 3 0 118 1 0 9 1 98 US300FT C 2 0...

Страница 12: ...annel A P1 P1 Current output AR ARS Upstream transducer signal for channel A P2 P2 Current output when specifying two current outputs BV BVS Downstream transducer signal for channel B P3 P3 Frequency...

Отзывы: