Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E  July 30 2015-00

 

 

 

Changes to the section “7.2 Flow Totalizers” (page 43) : 

Two of the key operations have been changed as below, requiring pressing the keys “three times” . 

To reset the two flow totalizers to zero:

 

Press key 

ON

O

  three times when a totalizer is displayed. 

To deactivate flow totalizing:

 

Press key 

OFF

O

 

three times when a totalizer is displayed. 

 

Addition to the section “14.5 Activation of a Pulse Output” (page 95) : 

Maximum pulse output rate is 2 pulse per second (+/-20%). 

 

Set the “Pulse Value” and the “Pulse Width” so that the actual maximum flow in the pipe is less than half   

the displayed value (“Max – Value”). 

Example: 

The actual maximum flow    120m

3

/h 

      Pulse Value = 0.02m

3

 

      Pulse Width = 150ms 

 

            INFO: Max – Value 

         240.0m

3

/h 

 

Attention! 

It is necessary to activate flow totalizers by pressing key 

ON

O

 to get the actual pulse 

output. Refer to the section 7.2 (page 43) for flow totalizers. 

 

 

Changes to the chapter “A Standard Specification” (page 109 to 116) : 

         Please use the following sheets for this chapter. 

 

Содержание US300FM

Страница 1: ...ite White Perform soldering Finish the brown cables as well as white cables You do not connect the thin lines to any terminals Brown 12 2 Less than 10mm 12 2 Perform soldering Shield line Cable shield...

Страница 2: ...ccurs A Tokuchu request is necessary for Submersible type However the diameter must be over 100 mm Changes to the section 3 5 5 Connection to the Power Supply page 20 AC power supply DC power supply T...

Страница 3: ...y this behavior Addition and deletion to the section 4 1 1 Key Operations page 23 and related pages Refer to the following table for the function of corresponding key operations The correct explanatio...

Страница 4: ...a wall thickness probe or some other ways and avoid mounting the transducers on such parts Attention Instructions for using Acoustic couplant weatherproof type Shin Etsu Silicone 1 COMPONENT RTV are a...

Страница 5: ...the section 14 5 Activation of a Pulse Output page 95 Maximum pulse output rate is 2 pulse per second 20 Set the Pulse Value and the Pulse Width so that the actual maximum flow in the pipe is less th...

Страница 6: ...Measuring Input One or two dual Channel A Channel B Current Output One or two Range 4 to 20 mA Load resistance 0 to 500 Note The current outputs may temporarily turn unstable during the power on seque...

Страница 7: ...tion only for test service maintenance purposes Time Programmable Measurement Function Automatic start stop of the measurement without human operation using internal clock for specific systems Can be...

Страница 8: ...suring data set in graphic format Marker type and color for each line of values selectable Scales for time axis and value axis can be changed from default condition of automatic scaling Graph printing...

Страница 9: ...th 4 4 m 14 4 ft cable CH Large High with 4 4 m 14 4 ft cable DG Very large General with 12 m 39 4 ft cable N Always N Fixing band strap and clips 6 B For 25 to 1 400 mm 1 to 55 in Two fixing bands On...

Страница 10: ...g fixture short ruler type for transducers medium pipe size type temperature range 30 to 200 C 22 to 392 F set of two blocks with 120 mm 4 7 in ruler USPA054 Mounting fixture standard type for transdu...

Страница 11: ...flowmeter US300FM Unit mm inch Transducers US300FT B Transducers US300FT C US300FT D Transducers X m inch Y m inch X Y m inch Weight kg lb US300FT B 2 0 78 7 1 0 39 4 3 0 118 1 0 9 1 98 US300FT C 2 0...

Страница 12: ...annel A P1 P1 Current output AR ARS Upstream transducer signal for channel A P2 P2 Current output when specifying two current outputs BV BVS Downstream transducer signal for channel B P3 P3 Frequency...