Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Addition to the section “3.5.5 Connection to the Power Supply” (page 20) :
Install an external switch or circuit breaker as a means to turn the power off (capacitance; 15A, conforming to
IEC947-1 and IEC947-3). Locate this switch either near the instrument or in other places facilitating easy
operation. Affix a “Power Off Equipment” label to this external switch or circuit breaker.
Changes to the section “3.6 Sensor ROM” (page 22) :
The sensor ROMs are packed and affixed to the transducers when shipped from the factory, not inserted to the
terminals in the main unit. They shall be connected to the corresponding terminals in the main unit when
installation.
Addition to the section “3.7 Replacement of the Fuse” (page 22) :
Be sure to use the fuse which satisfies the specification below.
Components may otherwise be damaged.
Changes to the section “3.7 Replacement of the Fuse” (page 22) :
Change the fuse specification “250V 1.25A, delayed action Type T” to as below.
“250V 1A delay type” for AC power supply
“250V 1.6A delay type” for DC power supply
Addition to the chapter “4. Getting Started” (page 23) :
Attention!
The current outputs may temporarily turn unstable during the power-on sequence
(including re-starting after power failure) and parameter display / setting mode. Take care
of your process not to be affected by this behavior.
Addition and deletion to the section “4.1.1 Key Operations” (page 23) and related pages :
Refer to the following table for the function of corresponding key operations.
The correct explanations for these key operations in the related pages are also as the table below.
Related pages:
“4.1.1 Interruption of Power Supply” (page 27)
“15.2 US300FM doesn’t react anymore” (page 106)
LIGHT
Switches the background lighting ON/OFF.
RESET
INIT
BRK
ENTER
[Operation during the measurement or menu display]
RESET: Press these three keys simultaneously to recover from an error. This
has the same effect as restarting the unit. Data will not be affected.
INIT (cold start): Press these three keys simultaneously and release
ENTER
first. After acknowledging the display of main menu, release (i)
BRK
key
first and then (ii)
C
key. This will initialize the instrument. Most
parameters and settings are reset to the factory default values. The
memory will not be cleared.
Note that when the data logging function is activated, the
DELETE
MEAS.VAL.
display will appear instead of main menu. In this case, after
releasing (i)
BRK
key first and then (ii)
C
key, select
NO
or
YES
and then
press
ENTER
to finish the procedure.
INIT
BRK
[Operation when powering the instrument on]
INIT (coldstart): Pressing these two keys simultaneously and after acknowledging the
display of main menu, release (i)
BRK
first and then (ii)
C
. This will initialize the
instrument. Most parameters and settings are reset to the factory default values.
The same procedure is required as above when the
DELETE MEAS.VAL.
display
appears.