Manual Change for IM 01G05B03-01E 1st Edition : No.15-010-E July 30 2015-00
Changes to the section “3.1 Scope of Delivery” (page 11) :
Change the figure of a nameplate to as below.
Addition to the section “3.1 Scope of Delivery” (page 11) :
When ordering a couplant weatherproof type (by couplant code “R” in the transducer specification or model code
USPA097), a package of “Shin-Etsu Silicone, 1 COMPONENT RTV (KE45T)” is delivered.
Addition to the section “3.3 General Precautions” (page 12) :
・
Do not disassemble or remodel.
Addition to the section “3.5.1 Location” (page 13) :
Attention!
The instrument must not be installed in a corrosive atmosphere.
Attention!
Use Submersible(IP68) type transducers when they are installed where condensation
occurs. A Tokuchu request is necessary for Submersible type. However the diameter
must be over 100 mm.
Changes to the section “3.5.5 Connection to the Power Supply” (page 20) :
AC power supply
DC power supply
Terminal Connection
Terminal
Connection
PE Earth
PE (Earth)
N(-) Neutral
N(-) -
DC
L(+)
Phase 100...230 V AC
±
10%
L(+)
+ DC 24V DC
±
10%
The terminal names are the
same both for AC and DC
power supply.
This can be used as the
earth terminal instead of
the terminal “PE”.
N(-)
PE
L(
+
)
KL3
(1) Terminal Block (page 19)
The outer shapes of the terminal block for
power supply connection, the printed
terminal names, and the design on the
board below the terminal block have been
changed as the drawing in the right.
(2) Terminal Assignment (page 21)
Regarding the terminal assignment, the
terminal names were revised as below
from the original manual. The needed
power supply is given in the nameplate on
the housing, not in the plate below the
terminal block.