7. BASIC SETTINGS
7.4 How to Set the Chart Speed
The chart speed can be set within a range of 1 to 12000 mm/h, depending on your
model. The initial value is 25 mm/h. To set the chart speed, proceed as follows:
MENU:
SETTING PROCEDURE:
Press the
MENU
-key for three seconds to enter the SET mode.
Select the ‘
SET=CHART
’ display by using the
UP/DOWN
-keys.
Then press the
ENT
-key.
In case of the pen model, select the desired chart speed using the
UP/DOWN
-keys, and press the
ENT
-key. The different possible
chart speeds are listed below in table 7.4.
In case of the dot model, set the desired chart speed using the
UP/
DOWN
and
RIGHT
-keys, and press the
ENT
-key.
Check in table 7.5 that all the printouts will be printed as expected.
Table 7.4 Chart Speeds in mm/h (pen model)
Table 7.5 Chart Speeds for which Printouts will not be performed
The setting is completed.
Press the
ENT
-key to return to the ‘
SPEED=1000 mm/h
’
display; or press the
ESC
-key to go to the ‘
SET=CHART
’
display; or press the
MENU
-key for three seconds to return to the
Operation mode.
NOTE
Setting the chart speed which goes into effect following a remote control signal (/R1
option), or following an alarm occurrence (/A1, 2 or 3 /R1 option), is
performed under the AUX-settings. See 8.3.
If the chart speed is high (600 to 1500 mm/h) in case of the pen model, the time tick in
the periodic printout will contain an error (e.g. if speed is 1500 mm/h, error will be
approx. 2cm).
If you selected a chart speed of 1 to 5 mm/h, in case of the dot model, select AUTO for
the trend recording format to prevent chart paper damage (see 8.1.1).
SPEED
1000
RANGE
ALARM
UNIT
CHART
CLOCK
COPY
AUX
7.4 How to Set the Chart Speed
SET=
CHART
MENU
ESC
ALARM
ACK
SPEED=
1000
ALARM
ACK
SET OK
*
*
3 sec
3 sec
MENU
ESC
mm/h
5
24
60
135
320
720
1500
3600
10800
6
25
64
150
360
750
1600
4000
12000
8
30
72
160
375
800
1800
4320
9
32
75
180
400
900
2000
4500
10
36
80
200
450
960
2160
4800
12
40
90
225
480
1000
2250
5400
15
45
96
240
500
1080
2400
6000
16
48
100
250
540
1200
2700
7200
18
50
120
270
600
1350
2880
8000
20
54
125
300
675
1440
3000
9000
Model
Alarm Printout/Recording Start/
Message Printout/Chart Speed Change/
Channel/Tag (dot model only)
pen
6/12 dot model
18/24 dot model
5 to 9 mm/h
1 to 9 mm/h, 101 to1500 mm/h
1 to 9 mm/h, 51 to1500 mm/h
Periodic Printout
1600 to 12000 mm/h
100 to 1500 mm/h
51 to1500 mm/h