IM 12B6B3-E-H
6-4 Calibration
pH
NO
YES
pH
NO
YES
YES
NO
NO
YES
pH
YES
pH
pH
NO
YES
ENT
pH
pH
pH
pH
NO
YES
Put sensors in buffer
solution. Press YES.
Set the value
using the >, , ENT key.
Select the flashing digit with the > key.
Increase its value by pressing the key.
When the correct value is displayed, press ENT to enter the change.
>
>
MEASURE
MAN.CAL
DISPLAY
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
TEMPERATURE
AUT.CAL
pH
CONTACTS
S1
S2
WASH/S3
FAIL/S4
pH
NO
YES
Manual Calibration. (2nd parameter calibration)
Press the MODE key. The legend AUT.CAL appears,
and the YES/NO key prompt flags flash. Press
NO.
The display MAN.CAL appears.
Press YES to start calibration.
Press YES or NO at NEW.SNS prompt.
YES
NO
For 2 point (As Pot and Slope)
Adjustment select second buffer
solution and adjust as for pH7 buffer.
NO
ENT
ENT
YES
WAIT is displayed
briefly then EXA returns
to measuring mode.