Parameter setting 5-1
IM 12B6B3-E-H
5. PARAMETER SETTING
5-1. Maintenance mode
Standard operation of the EXA instrument involves use of the maintenance (or operating) mode to set up
some of the parameters.
Access to the maintenance mode is available via the six keys that can be pressed through the flexible win-
dow in the instrument cover. Press the MODE-key once to enter this dialog mode.
NOTE:
At this stage the user will be prompted for pass code where this has been previously set up in service code
52 in chapter 5.
Automatic calibration
See “calibration” section 6.
Manual calibration
See “calibration” section 6.
Sample calibration
See “calibration” section 6.
Display setting
See “operation” section 4.
Manual impedance check
See “operation” section 5.
Setpoint
Select and adjust setpoint (when enabled in service menu section 5, ser-
vice code 51). See adjustment procedure in §5-1-5 and §5-2-1.
Wash
Manually start/stop wash cleaning (when enabled in service menu section
5, service code 51). See adjustment procedure in §5-1-4 and §5-2-4.
Temperature
Set automatic or manual compensation and adjust manual reading (when
pH is set in section 5 service code 01). See adjustment procedure in
§5-1-1.
Hold
Manually switch on/off HOLD (when enabled in commissioning menu sec-
tion). See adjustment procedure in §5-2-3..
Содержание PH402G
Страница 1: ...Instruction Manual Model PH402G pH Converter IM 12B6B3 E H 7th Edition YOKOGAWA...
Страница 2: ......
Страница 6: ...IM 12B6B3 E H...
Страница 26: ...IM 12B6B3 E H...
Страница 49: ...Parameter setting 5 17 IM 12B6B3 E H 5 3 Notes for guidance in the use of service coded settings...
Страница 68: ...IM 12B6B3 E H...
Страница 76: ...IM 12B6B3 E H...
Страница 82: ...IM 12B6B3 E H...
Страница 93: ...IM 12B6B3 E H...