IM 12B6B3-E-H
6-2 Calibration
pH
pH
YES
NO
YES
NO
YES
NO
pH
YES
NO
pH
NO
pH
YES
Press YES for single point
(As Pot) adjustment.
Press NO to proceed to calibrate Slope.
The instruments waits for
the reading to stabilize.
(The pH display flashes)
When reading Is stable,
the CAL END message
appears.
Transfer to second buffer (pH4) and press YES.
After briefly displaying WAIT, the display
returns to the normal readout.
The display now flashes the
pH value. The instrument
waits for the reading to stabilize.
NO
YES
pH
NO
YES
pH
YES
MEASURE
MAN.CAL
DISPLAY
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
TEMPERATURE
AUT.CAL
pH
CONTACTS
S1
S2
WASH/S3
FAIL/S4
NO
YES
pH
Automatic Calibration
Press the MODE key. AUT.CAL appears in the display, and the
YES/NO key prompt flags flash. Press YES.
NEW.SNS: Display flashes YES/NO
answer YES if new sensor fitted or NO if not.
Care! YES resets logbook calibration data.
Insert the sensors in pH 7 buffer solution.
Press YES to start calibration.
NO
Note: To start calibration with another solution, press NO to cycle through the other options.
"CAL 7" NO "CAL 4" NO "CAL 9" NO "CAL 7"
YES
When the reading is stable, the CAL END
message appears. Press YES.
After briefly displaying WAIT, the display returns to
the normal readout.
6-5. Calibration procedures
6-5-1. Automatic calibration
Содержание PH402G
Страница 1: ...Instruction Manual Model PH402G pH Converter IM 12B6B3 E H 7th Edition YOKOGAWA...
Страница 2: ......
Страница 6: ...IM 12B6B3 E H...
Страница 26: ...IM 12B6B3 E H...
Страница 49: ...Parameter setting 5 17 IM 12B6B3 E H 5 3 Notes for guidance in the use of service coded settings...
Страница 68: ...IM 12B6B3 E H...
Страница 76: ...IM 12B6B3 E H...
Страница 82: ...IM 12B6B3 E H...
Страница 93: ...IM 12B6B3 E H...