Parameter setting 5-25
IM 12B6B3-E-H
Code Display
Function
Function detail
X
Y
Z
Default Values
mA Outputs
30
*mA
output signal
0-20 mA on mA1
0
1.1
4-20 mA on mA1
1
mA1 4-20
0-20 mA on mA2
0
4-20 mA on mA2
1
mA2 4-20
31
The choices below appear when pH was selected in service code 01
pH
*OUTP.F
Output Parameter
pH on mA1
0
0.2 pH on
mA1
pH (table) on mA1
1
Parameter 2 (ORP or rH) on mA1
2
pH on mA2
0
pH (table) on mA2
1
Temperature on mA2
2
Temp. on
mA2
pH PI control on mA2
3
Parameter PI control on mA2
4
*D/R
Direct or reverse
For rising signal, set
0
0
Reverse
acting PI control
for decreasing mA output
For rising signal, set
1
for increasing mA output
31
The choices below appear when ORP was selected in service code 01
ORP
*OUTPUT.F
Output parameter
ORP on mA1
0
0.2 ORP on
ORP (table) in mA1
1
mA1
ORP on mA2
0
ORP (table) on mA2
1
Temperature on mA2
2
Temp on
ORP PI control on mA2
3
mA2
*D/R
Direct or reverse
For rising signal, set
0
0
Reverse
acting PI control
for decreasing mA output
For rising signal, set
1
for increasing mA output
32
*BURN
Burn function
mA 1 No burnout
0
0.0 No Burn.
mA 1 Burnout downscale
1
mA 1 Burnout upscale
2
mA 2 No burnout
0
No Burn.
mA 2 Burnout downscale
1
mA 2 Burnout upscale
2
33
*RG.mA2
PI range
Proportional range for mA control signal
*pH.mA2
Display depends on
(use >, ^, ENT keys to adjust value)
1.0 pH
*mV.mA2
parameter selected
2.0 rH
*rH.mA2
in code 31
50 mV
34
*tI.mA2
Integral time
100 sec.
(for PI control)
35
*TABL1
Output table for mA1
*0%
Linearisation table for mA1 in 5% steps.
*5%
The measured value is set in the main
*10%
display using the >, ^, ENT keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value may
*90%
be skipped, and a linear interpolation will
*100%
take place.
36
*TABL2
Output table for mA2
*0%
Linearisation table for mA2 in 5% steps.
*5%
The measured value is set in the main
*10%
display using the >, ^, ENT keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value may
*90%
be skipped, and a linear interpolation will
*100%
take place.
37-39
Not used
Содержание PH402G
Страница 1: ...Instruction Manual Model PH402G pH Converter IM 12B6B3 E H 7th Edition YOKOGAWA...
Страница 2: ......
Страница 6: ...IM 12B6B3 E H...
Страница 26: ...IM 12B6B3 E H...
Страница 49: ...Parameter setting 5 17 IM 12B6B3 E H 5 3 Notes for guidance in the use of service coded settings...
Страница 68: ...IM 12B6B3 E H...
Страница 76: ...IM 12B6B3 E H...
Страница 82: ...IM 12B6B3 E H...
Страница 93: ...IM 12B6B3 E H...