IM 04P01B01-01E
5-12
5.3 Setting the Unit on Scaled Channels
Units are set on channels whose input range is set to Scaling, 1-5V, or SQRT.
Procedure
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the
key or
SHIFT
+
key to select
Unit
and then press the key.
3.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
4.
When the
Setting complete
screen appears, do either of the following:
Press the key to set other channels.
If you are done, press the
ESC
key.
5.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Set the unit.
Set the channel range.
The new setting takes effect.
First channel
Last channel
ESC/?
Set=Unit
CH=01-01
Unit=
01-01 Channel
Setting complete
Explanation
Characters That Can Be Used for Units
A unit is set using up to six characters.
The available characters are as follows:
Alphabet, numbers, symbols (%, #, °, @, +, –, *, /, (, ), µ, Ω,
2
,
3
, .), and space
<Related Topics>
Entering the unit string: Page 25 in the μR10000 Recorder Operation