5-5
IM 04P01B01-01E
Frequently Used Setup Operations (Setting Mode)
5.1 Setting the Input Range
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
ON/OFF Input
Select DI.
Set the channel range.
Select Cont or Level.
Set the left span value.
Set the right span value.
The new setting takes effect.
First channel Last channel
ESC/?
Set=Range
CH=01-01
MOde=DI
Range=Level
Span_L=0
Span_R=1
01-01 Channel
Setting complete
Description
• Selectable Range Type, Span Left, and Span Right
The range type, span left, and span right can be set in the range shown below. Span
left and span right cannot be set to the same value.
Range Type
Selectable Span Values
Level
0: Less than 2.4 V, 1: 2.4 V or higher, within ±6 V
Contact
0: Open, 1: Closed