5.1 Setting the Input Range
Input range is set for each measurement channel. Set unused channels to Skip.
• The input range cannot be changed on models with the computation function (/M1
option) when computation is in progress.
• If you change the input range, set the bias, alarm, partial expanded recording, and
calibration correction again.
Procedure
TC, RTD, and DC Voltage
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Press the key with
Range
shown on the screen.
3.
Set each item and press the key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
4.
When the
Setting complete
screen appears, do either of the following:
Press the key to set other channels.
If you are done, press the
ESC
key.
5.
Hold down the
MENU
key for 3 seconds to return to Operation mode.
Select TC, RTD, or Volt.
Set the channel range.
Select the range type.
Select the left span value.
Select the right span value.
The new setting takes effect.
First channel Last channel
ESC/?
Set=Range
CH=01-01
Mode=TC
Range=R
Span_L=
0.0
Span_R= 1760.0
01-01 Channel
Setting complete