<Start-up and Safety Precautions>
9
IM 12F05B10-01EN 2nd Edition : June 01, 2021-00
2. Wiring and Installation
For details, refer to the User’s Manual IM 12F05B10-02EN.
2.1
Installation Site
Select an installation site where
Ambient Operating temperature:
-10 to +55 ºC
Storage Temperature:
-30 to +70 ºC
Humidity:
10 to 90% RH at 40ºC (Non-condensing)
Install the FC800D Non-reagent type Free Chlorine Sensor Unit at a location that
● is free from rain water such as an indoor location or in a cabinet.
● has low vibration.
● has low corrosive gas.
● is low humidity.
● has low temperature variation and where the temperature is maintained at or as near to
room temperature as possible.
● allows for sufficient maintenance space with easy maintenance access.
● allows for drai
nage.
For information on FLXA402T, read the user’s manual.
Install the FC800D’s converter and sensor unit in pipes, racks or walls using their special
mounting brackets, respectively. Note that these mounting brackets are supplied only when
specified. Refer to General Specifications (GS 12F05B10-01EN) for details.
n
Piping without Sampling System
When sample has a pressure of 1 to 150 kPa and a flow rate of 0.1 to 2.5 L/min, such sample can
be directly introduced into the measuring tank. Connect a Ø6 X Ø4 mm polyethylene tube to the
sample water inlet.
Connect a Ø22 X Ø15 mm flexible mesh-reinforced tube to the drain outlet from the measuring
tank. Install this tube so that no water entrapment is formed. Entrapment of water in the drain
tube may cause overflow from the measuring tank.