<Start-up and Safety Precautions>
11
IM 12F05B10-01EN 2nd Edition : June 01, 2021-00
2.2 Wiring
WARNING
When Standard “-AD” (General purpose for CSA) with conduit adapter “/C□3” is specified, always
use a dedicated plate for wiring to comply with safety regulations. For Option “/TB” the plate is
supplied with product. Use the plate with Terminal Box to provide wiring.
When wiring FC800D, first remove the cover of the smart box, then attach the cable gland.
After the wiring completes, install the shield cover securely. Refer to user’s manual IM 12F05B10-
02EN for details.
(201)L
(202)N
(002)G
L
N
G
(32)NC
(31)C
(33)NO
S1
(42)NC
(41)C
(43)NO
S2
(72)NC
(71)C
(73)NO
S4
(61) +
(62) -
mA1
(65) +
(66) -
mA2
(81) +
(82) -
mA3
(85) +
(86) -
mA4
(21)
(22)
(24)
SHLD(63)
(87) +
(88) -
SHLD(89)
A+(91)
B-(92)
GND(93)
SHLD(94)
Termination(95)
L(101)
N(102)
G(001)
(52)NC
(51)C
(53)NO
S3
Power
Converter
Power
2nd input
(-FC, -TB, -P, -C)
*2
*2
SENCOM
1st input
(82)SHLD
(86)GND
(83)B-
(84)A+
(87)V5
*1: Power terminal “G” on Terminal Box must be grounded (ground resistance: 100 ohm or less).
*2: Connection cable is normally 1 m in length. It is avaiable with /L02, /L03, /L05, /L10 or /L20 depending on the length you need.
FC800D
Non-reagent Type Free Available
Chlorine Sensor Unit
Power
Supply
*1
FLXA402T
Liquid Analyzer for Turbidity and Chlorine
Analog output (-N2, -N4)
Analog output (-N2, -N4)
Analog output (-N4)
Analog output (-N4)
Contact input (-N2, -N4)
Analog input (-N4)
RS485 (-R)
Ethernet (-E)
Contact output
(-WR)