5-13
IM 01F06A00-01E
5. PARAMETER SETUP
(5) Item E (METER SETUP)
These items are for detector set up that has already been set
before shipment.
A value in “( )” is the data corresponding to indicator.
[E10:NOMINAL SIZE] Nominal size of the
detector
Select the nominal size of the flowmeter, from “15mm(0),
25mm(1), 40mm(2), 50mm(3), 80mm(4), 100mm(5),
150mm(6), 200mm(7), 250mm(8), 300mm(9)”
[E20:BODY TYPE] Body type for the detector
Select body type for detector from standard or high pressure.
Standard
(0) : Standard type
High Pressure
(1) : High Pressure type (tokuchu)
Low Flow Unit (1)
(2) : Reduced Bore type (option : /R1)
Low Flow Unit (2)
(3) : Reduced Bore type (option : /R2)
NOTE
Parameter setting for the Reduced Bore type,
Select Low Flow Unit (1) or (2) and set. Set
nominal size of the model code to E10:
NOMINAL SIZE.
[E30:SENSOR TYPE] Sensor type for the
detector
Select sensor type for the detector.
Standard
(0) : Standard type
High Temperature (1) : Option /HT
Low Temperature (2) : Option /LT
[E40:K-FACTOR UNIT],
Select this unit from p/l, p/Usgal, p/Ukgal.
IMPORTANT
K-FACTOR is the eigenvalue of each detector.
Please keep the factory preset value.
NEVER REWRITE IT.
(Unless the replacement of the remote type detector.)
[E41:K-FACTOR]
The flowmeter data plate includes a K-factor (KM) at 15
°
C
for the combined detector.
[E50:DETECTOR NO.] Detector number of
flowmeter
Set the serial number using 16 alphanumeric characters of the
detector combined converter.
(6) ltem F (Thermometer)
These items are for setting of thermometer and available
when build in thermometer type (Option code: /MV).
[F10: Function] Thermometer function
Select the thermometer function.
Monitor only(0):
Only temperature measurement.
Saturated Steam:
Mass Flow rate is calculated from density
values by temperature measurement
using saturated steam table.
Superheat Steam: Mass Flow rate is calculated from density
values by temperature measured by using
steam table. In order to measure
superheat steam. It is necessary to make
constant pressure value.
GAS: STD/Normal: Volumetric flow rate at standard
condition is calculated by using Pressure-
Temperature correction. It is necessary to
male constant pressure value.
Liquid: Mass:
Mass flow rate is calculated by using the
density change values depend on
temperature values by which the
secondary order function is used.
The following item should be done in case of
which F10 is Saturated steam
[F12 MASS UNlT] Mass flow unit
Select mass rate unit from kg(0), t(1), lb(2), k lb(3).
The following items should be done in case of
which F10 is Superheat steam
[F14 PRESS UNIT] Pressure unit
Select pressure unit from MPa abs(0), kPa abs(1), bar abs(2),
kg/cm
2
a(3), psia(4).
[F15 PRESS f] Pressure value
Set absolute pressure values at operating condition.
[F16 MASS UNIT] Mass flow unit
Select mass flow unit from kg(0), t(1), lb(2), k lb(3).
The following items should be done in case of which F10 is
GAS: STD/Normal
[F18 TEMP UNIT] Temperature unit
Select temperature unit from deg C(0), deg F (1).
[F19 TEMP b] Temperature b
Set temperature value at normal/standard condition.