5-10
IM 01F06A00-01E
5. PARAMETER SETUP
[B22:100% FREQ] Pulse numbers of 100% at
one second
Set pulse number at 100% for one second when “FRE-
QUENCY” in B20 is selected.
F050401.EPS
Pulse numbers (PPS)
T=1s
[B23:SET LEVEL] Level of flow switch
Set level of flow switch when “FLOW SW” in B20 is
selected. The contact output is sent out when the flowrate is
less than the set comparison level.
Output
(%)
Setting Value
10% Hysteresis
Contact Output ON : FLOW SW (LOW : ON)
Contact Output OFF : FLOW SW (LOW : OFF)
F050402.EPS
[B30:UPPER DISP] Upper indicator display
Select upper display, Flow rate (%) (0), Flowrate (1),
TEMP(%)(2). “TEMP(%)” can be selected when Option
Code
/MV.
[B31:LOWER DISP] Lower indicator display
Select lower indicator display, “BLANK (0), TOTAL (1),
TEMP(2). When “BLANK” in B31 is selected, indicator is
blank. “TEMP” can be selected when Option Code
/MV.
[B40:TOTAL START]
Select the START/STOP of totalizer from “STOP (0),
START (1).”
[B45:TOTAL RATE] Total rate of the totalizer
Set the total rate of the totalizer.
[B47:TOTAL RESET] Reset the totalizer
When totalizer reset function is executed, the total display
and communication parameter are reset.
The following items should be done in case of
which Option code “/MV” is selected.
[B50 A/OUT SELECT] Analog Output select
Select the analog output select from flow rate or temperature.
When changing the analog output, UPPER DISPLAY can be
changed shown as below automatically.
B50 : A/OUT SELECT
“TEMP” TO “FLOW”
“FLOW” TO “TEMP”
UPPER DISPLAY
FLOW(%)
TEMP(%)
(“B30 : UPPER DISPLAY” is “FLOW RATE”, it can not be changed.)
T050402.EPS
The following item should be done in case of
which B50 is “TEMP”
[B51 TEMP 0%]
Set temperature value of 0% output.
[B52 TEMP 100%]
Set temperature valve of 100% output.