<5. Parameter Summary>
5-4
IM 01C25T01-06EN
Function
Label
Item
Contents
Default value Handling
*
1
Diag Variables
EJX
Ratio fDP
SQRT (fDP/Ref fDP).
RG
Ratio fDP Status
Status of Ratio fDP
RG
Ratio fSPl
SQRT (fSPl/Ref fSPl).
RG
Ratio fSPl Status
Status of Ratio fSPl
RG
Ratio fSPh
SQRT (fSPh/Ref fSPh).
RG
Ratio fSPh Status
Status of Ratio fSPh
RG
BlkF
Blockage degree characterized in
comparison of high-pressure side
and low-pressure side pressure
fluctuation values
RG
BlkF Status
Status of BlkF
RG
DPAvg
Ratio of the average of differential
pressure/pressure to the maximum
span of an EJX transmitter.
RG
DPAvg Status
Status of DPAvg
RG
CRatio fDP
Compensated fDP
RG
CRatio fDP Status
Status of CRatio fDP
RG
NRatio fDP
Non-compensated fDP
RG
NRatio fDP Status
Status of NRatio fDP
RG
Display setup Bar Indicator
Bar indicator
Off or On
On
W
Chg power on info
Change the LCD display when
powering on
On or Off
M
Disp Out 1
LCD output 1
PRES, PRES %,
ENGR.PRES, SP, SP %
PRES %
W
Disp Out 2
LCD output 2
PRES, PRES %,
ENGR.PRES, SP, SP % or
Not used
Not used
W
Disp Out 3
LCD output 3
(Ditto)
Not used
W
Disp Out 4
LCD output 4
(Ditto)
Not used
W
Disp Pres % fnctn
% display mode
Linear or Sq root
As specified or
Linear
W
Disp
Pres
% Reso
% display resolution
Normal or High resolution
Normal
W
Engr exp
User set exponent
---, x10, x100, x1000
--- or as
specified
W
Engr LRV
User set lower range value
Unit specified in Set Engr Unit As specified
W
Engr point
Decimal place for user set
0 to 4
2
W
Engr URV
User set upper range value
Unit specified in Set Engr Unit As specified
W
Modify Engr Unit
User set engineering unit
M
Pres disp point
Decimal place for differential
pressure
0 to 4
2
W
Set Engr Unit
Engineering unit select
M
SP disp point
Decimal place for static pressure
0 to 4
2
WD
Squawk
HART 7
Squawk
Specify the device under the
communication (turn the LCD
on).
M
Error log
Error log Clear
Clear error records
M
Error log view
Error records
Log1 (latest) to log4
M
*1: Handling:
R
=Read only,
W
=Read & Write,
M
=Method,
A
=Applicable for option code AL,
G
=Applicable for option code DG6,
D
=Applicable for differential pressure transmitters. Do not change these parameters for pressure transmitters.
*2:
Max three Burst Messages (Burst Message 1, Burst Message 2, Burst Message 3)
*3:
The default value shows MWP (Maximum working pressure) of the capsule.
Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.
*4:
For output signal code J, refer to the section 3.3.18.
*5: These parameters may contain adjustment values based on the customer range calibration at the factory upon shipment.
When executing “Clear P sensor trim”, P LTD and P UTD will become “0”, and P LTP and P UTP will become the value of LRV and
URV respectively.