# Parameter name Default value
Explanation
28 Ref fSPl
Value of
fSPl
obtained under normal condition.
29 Ref fSPl Status
Status of
fSPl
obtained under normal condition.
30 Ref fSPh
Value of
fSPh
obtained under normal condition.
31 Ref fSPh Status
Status of
fSPh
obtained under normal condition.
32 Ref BlkF
Value of
BlkF
obtained under normal condition.
33 Ref BlkF Status
Status of
BlkF
obtained under normal condition.
34 Ref DPAvg
Value of
DPAvg
obtained under normal condition.
35 Ref DPAvg
Status
Status of
DPAvg
obtained under normal condition.
36 Lim fDPmax
Refer to
Table 4.3
Upper limit for
Ratio fDP
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
37 Lim fDPmin
Lower limit for
Ratio fDP
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
38 Lim fSPlmax
Upper limit for
Ratio fSPl
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
39 Lim fSPlmin
Lower limit for
Ratio fSPl
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
40 Lim fSPhmax
Upper limit for
Ratio fSPh
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
41 Lim fSPhmin
Lower limit for
Ratio fSPh
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
42 Lim BlkFmax
Upper limit for
BlkF
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
43 Lim BlkFmin
Lower limit for
BlkF
to detect the blockage.
Writable only when
Diag Mode
is "Stop".
44 Lim DPAvgmax
Upper limit for
DPAvg
. Writable only when
Diag Mode
is "Stop".
45 Lim DPAvgmin
Lower limit for
DPAvg
. Writable only when
Diag Mode
is "Stop".
46 Ref Lim fDPmin
7.0E-10
Lower limit to judge whether
Ref fDP
is available for ILBD operation.
Writable only when Diag Mode is "Stop".
47 Ref Lim fSPmin
1.0E-10
Lower limit to judge whether
Ref fSPl
and
Ref fSPh
are available for ILBD
operation.
Writable only when Diag Mode is "Stop".
48 Ref Lim BlkFmax 0.5
Upper limit to judge whether
Ref BlkF
is available for ILBD operation.
Writable only when Diag Mode is "Stop".
49 Status group 8
Refer to Table 4.5.
50 Status group 9
Refer to Table 4.5.
51 CRatio fDP
Ratio fDP
is compensated by following formula and used as treatable monitoring
value when the flow change is too large or small.
Sqrt (
fDP / Ref fDP
) X |
Ref DPAvg / DPAvg
|
When compensation is selected in
Diag DP Comp
,
CRatio fDP
is used as
monitoring value.
52 CRatio fDP
Status
Status of
CRatio fDP
53 NRatio fDP
When Non-compensation is selected in
Diag DP Comp
, NRatio fDP is used as
monitoring value.
NRatio fDP
= Sqrt (
fDP / Ref fDP
)
54 NRatio fDP
Status
Status of
NRatio fDP
55 Diag DPComp
0:
Compensation
Whether
fDP
is referred by
CRatio fDP
or
NRatio fDP
is selected.
56 Diag Applicable
After the reference value is obtained, the applicable blockage detection is
displayed on this parameter.