background image

4

Large capacity memory up to 250 Mpoints

Long memory is necessary to keep high speed sample rate in long term measurement.

<Basic Formula>  Measuring time = Memory length/Sample rate

If 250 Mpoints (Memory expansion option /M3) is installed, Max. 0.2 sec waveform can be captured even at 1.25 GS/s sample rate when 
taking 2 ch measurements in Single mode.

Zoom & search function

With 2 different zoom location at the same time and variety of search function lets you pull out and display necessary data effectively.

Zoom two locations simultaneously

Because the DLM2000 series lets you set zoom factors 
independently, you can display two zoomed waveforms with 
different time axis scales at the same time. Also, using the Auto 
Scroll function, you can automatically scroll waveforms captured 
in long memory and change the zoomed location. With Auto Scroll 
you can choose forward, backward, fast-forward, scroll speed, and 
other control options.

Waveform search using edge criterion

Zoom Search function

This function searches captured waveforms in the long memory 
and displays waveforms that meet the search criteria in the zoom 
area. The locations of the found waveforms are marked on screen 
( shows the current location).

• Waveform search criteria

Edge, edge (with conditions), state pattern, pulse width, state width, 
serial bus (only on models with the serial bus analysis option)

Detailed waveform measured for 20 seconds are 

shown in 20 milliseconds and 100 microseconds 

span.

Waveform in 250 Mpoints can be magnified 

up to × 200000.

apacity

Best-in-class long memory

C

Relationship between measuring time and sample rate in 250 Mpoint

Sample rate

Maximum measuring time

1.25 GS/s

0.2 s

125 MS/s

2 s

12.5 MS/s

20 s

1.25 MS/s

200 s

125 kS/s

2000 s

62.5 kS/s

5000 s

Search results 

marked

Auto Scroll menu

Zoom 1

Zoom 2

Caution is needed when using oscilloscope that does not have enough memory, which 
can cause lack of sample rate and possible failure capturing accurate waveform.

Sample rate is too low.

Sample rate is fairly high.

Zoom 1

×1000

Zoom 2

×200000

Содержание DLM2000 Series

Страница 1: ...s 200 MHz 350 MHz 500 MHz bandwidth models Lightweight and compact Large 8 4 inch LCD display Long memory Up to 250 M points with M3 option High speed sampling Up to 2 5 GS s 1 25 GS s with 4 ch DLM 2...

Страница 2: ...8 bit logic Example of logic probe connection Logic probe for the DLM2000 Large screen in a compact body Footprint is approximately 2 3 the size of an A4 size paper depth of approximately 200 mm Fast...

Страница 3: ...trol options Waveform search using edge criterion Zoom Search function This function searches captured waveforms in the long memory and displays waveforms that meet the search criteria in the zoom are...

Страница 4: ...nd other operations on history waveforms Using the History function you can analyze rarely occurring abnormal signals Searching for waveforms in zones created by moving measured waveforms up down left...

Страница 5: ...g observation of only the desired bandwidths Trigger Function capturing combined analog digital complex waveforms The DLM2000 series comes with a variety of easy to configure triggers combining analog...

Страница 6: ...risons Has a GO NO GO function Action on trigger GO NO GO can be determined using trigger conditions zone waveforms measurement parameters and other criteria For NO GO actions can be carried out at th...

Страница 7: ...2 C SPI UART LIN CAN CANFD FlexRay SENT PSI5 CXPI Analog input Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Logic input Yes Yes Yes NA NA NA NA Yes NA NA Simultaneous analyses of I2 C and SPI Four bus deco...

Страница 8: ...DLM2032 710115 350 MHz DLM2052 710125 500 MHz DLM2024 710110 200 MHz 4 analog channels or 3 analog channels 8 bit logic DLM2034 710120 350 MHz DLM2054 710130 500 MHz Analog Signal input Input channel...

Страница 9: ...mputable memory length M1 M1S 25 MPoints M2 62 5 MPoints M3 125 MPoints Reference function Up to 2 traces REF1 REF2 of saved waveform data can be displayed and analyzed Action on trigger Actions Buzze...

Страница 10: ...1 us to 100 us with resolution of 0 01 us Data type Fast channel Nibbles User Defined Slow channel Short Enhanced SENT trigger modes Start of fast channel Analyzable no of frames 100000 frames max Li...

Страница 11: ...s appearing in this document are the registered trademarks or trademarks of their respective companies Additional Option License for DLM2000 1 Model Suffix code Description 709810 G2 User defined math...

Отзывы: