10
Logic Signal Input (4 ch model only)
Number of inputs
8 bit (excl. 4 ch input and logic input)
Maximum toggle frequency
*1
Model 701988: 100 MHz, Model 701989: 250 MHz
Compatible probes
701988, 701989 (8 bit input) (701980, 701981 are available)
Min. input voltage
701988: 500 mVp-p, 701989: 300 mVp-p
Input range
Model 701988: ±40 V, Model 701989: threshold ±6 V
Max. nondestructive input voltage ±40 V (DC + ACpeak) or 28 Vrms (when using 701989)
Threshold level setting range
Model 701988: ±40 V (setting resolution of 0.05 V)
Model 701989: ±6 V (setting resolution of 0.05 V)
Input impedance
701988: Approx. 1 MΩ/approx. 10 pF
701989: Approx. 100 kΩ/approx. 3 pF
Maximum sampling rate
1.25 GS/s
Maximum record length (Points)
Repeat
Single
/M1
6.25 M
25 M
/M2
12.5 M
62.5 M
/M3
25 M
125 M
Triggers
Trigger modes
Auto, Auto Level, Normal, Single, N-Single
Trigger type, trigger source
A triggers
Edge
CH1 to CH4, Logic, EXT, LINE
Edge OR
CH1 to CH4
Edge Qualified CH1 to CH4, Logic, EXT
State
CH1 to CH4, Logic
Pulse width
CH1 to CH4, Logic, EXT
State width
CH1 to CH4, Logic
TV
CH1 to CH4
Serial Bus
I
2
C (optional)
CH1 to CH4, Logic
SPI (optional)
CH1 to CH4, Logic
UART (optional)
CH1 to CH4, Logic
FlexRay (optional) CH1 to CH4
CAN (optional)
CH1 to CH4
CAN FD (optional) CH1 to CH4
LIN (optional)
CH1 to CH4
SENT (optional)
CH1 to CH4, Logic
PSI5 (optional)
CH1 to CH4
User defined
CH1 to CH4
AB triggers
A Delay B
10 ns to 10 s (Edge, Edge Qualified, State, Serial Bus)
A to B(N)
1 to 10
9
(Edge, Edge Qualified, State, Serial Bus)
Dual Bus
Serial Bus only
Trigger level setting range
CH1 to CH4 ±4 div from center of screen
Trigger level setting resolution CH1 to CH4 0.01 div (TV trigger: 0.1 div)
Trigger level accuracy
*1
CH1 to CH4 ±(0.2 div + 10% of trigger level)
Window Comparator
Center/Width can be set on individual Channels from CH1 to CH4
Display
Display
*5
8.4-inch TFT color liquid crystal display, 1024 × 768 (XGA)
Functions
Waveform
acquisition modes
Normal, Envelope, Average
High Resolution
mode
Max. 12 bit (the resolution of the A/D converter can be improved
equivalently by placing a bandwidth limit on the input signal)
Sampling modes
Real time, interpolation, repetitive sampling
Accumulation
Select OFF, Intensity (waveform frequency by brightness), or Color
(waveform frequency by color)
Accumulation time: 100 ms to 100 s, Infinite
Roll mode
Enabled at 100 ms/div to 500 s/div (depending on the record length setting)
Zoom function
Two zooming windows can be set independently (Zoom1, Zoom2)
Zoom factor
×2 to 2.5 points/10 div (in zoom area)
Scroll
Auto Scroll
Search functions Edge, Edge Qualified, State, Pulse Width, State Width,
I
2
C (optional), SPI (optional), UART (optional), CAN
(optional), CAN FD (optional), LIN (optional), FlexRay
(optional), SENT (optional), PSI5 (optional), CXPI (optional)
History memory
Max. data (record length 1.25 k Points, with)
/M1 or /M1S: 10000, /M2: 20000, /M3: 50000
History search
Select Rect, Wave, Polygon, or Parameter mode
Replay function
Automatically displays the history waveforms sequentially
Display
Specified or average waveforms
Cursor
Types
∆T, ∆V, ∆T & ∆V, Marker, Degree
Snapshot
Currently displayed waveform can be retained on screen
Computation and Analysis Functions
Parameter
measurement
Max, Min, P-P, High, Low, Amplitude, Rms, Mean, Sdev, ,
IntegTY, +Over, −Over, Pulse Count, Edge Count, V1, V2, ∆T, Freq,
Period, Avg Freq, Avg Period, Burst, Rise, Fall, +Width, −Width, Duty,
Delay
Statistical computation
of parameters
Max, Min, Mean,
σ
, Count
Statistics modes
Continuous, Cycle, History
Trend/Histogram display
of wave parameters
Up to 2 trend or histogram display of specified wave parameters
Computations (MATH) +, −, ×, Filter (Delay, Moving Avg, IIR Lowpass, IIR Highpass), Integ,
Count (Edge, Rotary), user defined math (optional)
Computable no. of
traces
2 (Math1, Math2) (1 trace for 2 ch model)
Max. computable
memory length
/M1, /M1S: 25 MPoints, /M2: 62.5 MPoints, /M3: 125 MPoints
Reference function
Up to 2 traces (REF1/REF2) of saved waveform data can be displayed
and analyzed
Action-on-trigger
Actions: Buzzer, Print, Save, Mail
GO/NO-GO
Modes: Rect, Wave, Polygon, Parameter
Actions: Buzzer, Print, Save, Mail
XY
Displays XY1, to XY2 and T-Y simultaneously
FFT
Number of points: 1.25 k, 12.5 k, 25 k, 125 k, 250 k
Window functions: Rectangular, Hanning, Flat-Top
FFT Types: PS (LS, RS, PSD, CS, TF, CH are available with /G2 or /G4
option)
Histogram
Displays a histogram of acquired waveforms
User-defined math
*6
(/G2 and /G4 options)
The following operators can be arbitrarily combined in equations:
+, −, ×, /, SIN, COS, TAN, ASIN, ACOS, ATAN, INTEG, DIFF, ABS, SQRT,
LOG, EXP, LN, BIN,DELAY, P2 (power of 2), PH, DA, MEAN, HLBT,
PWHH, PWLL, PWHL, PWLH, PWXX, FV, DUTYH, DUTYL, FILT1, FILT2
The maximum record length that can be computed is the same as the
standard math functions.
Power supply analysis (/G3 and /G4 options)
*6
Power analysis
For Pwr1 and Pwr2, selectable from 4 analysis types
Deskweing between the voltage and current waveforms can be
executed automatically.
Switching loss Measurement of total loss and switching loss, power
waveform display, Automatic measurement and
statistical analysis of power analysis items (Wp, Wp+,
Wp−, Abs.Wp, P, P+, P−, Abs.P, Z)
Safety
operation area
SOA analysis by X-Y display, using voltage as X axis,
and current as Y axis is possible
Harmonic
analysis
Basic comparison is possible with following standard
Harmonic emission standard IEC61000-3-2 edition
2.2, EN61000-3-2 (2000), IEC61000-4-7 edition 2
Joule integral
Joule integral (I
2
t) waveform display, automatic
measurement and statistical analysis is possible
Power
Measurement
Automated measurement of power parameters for up to two pairs of
voltage and current waveforms. Values can be statistically processed
and calculated.
Measurement
parameters
Urms, Unm, Udc, Urmn, Uac, U+pk, U−pk, Up−p,
Irms, Imn, Idc, Irmn, Iac, I+pk, I−pk, Ip−p, P, S, Q, Z,
λ
,
Wp, Wp+, Wp−, Abs.Wp, q, q+, q−, Abs.q, Avg Freq
(voltage, current)
Common Features of Serial Bus Signal Analysis Functions (/F1 to /F11 Options)
Analysis result display
Decoded information is displayed together with waveforms or in
list form.
Auto setup function
A threshold value, time axis scale, voltage axis scale and other
bus-specific parameters such as a bit rate and recessive level are
automatically detected. Trigger conditions are set based on the
detected result and decoded information is displayed.
(The type of a bus signal needs to be specified in advance.)
Search function
Search of all waveforms for a position that matches a pattern or
condition specified by data information.
Analysis result saving
function
Analysis list data can be saved to CSV-format files. Trend data can
be also saved for SENT signals.
I
2
C Bus Signal Analysis Functions (/F2 and /F3 Options)
*6
Applicable bus
I
2
C bus Bus transfer rate: 3.4 Mbit/s max.
Address mode: 7 bit/10 bit
SM bus Complies with System Management Bus
Analyzable signals
CH1 to CH4, Logic input, or M1 to M2
I
2
C Trigger modes
Every Start, Address & Data, Non-Ack, General Call, Start Byte, HS
Mode
Analyzable no. of data 300000 bytes max.
List display items
Analysis no., time from trigger position (Time (ms)),1st byte address,
2nd byte address, R/W, Data, Presence/absence of ACK, information
SPI Bus Signal Analysis Functions (/F2 and /F3 Options)
*6
Trigger types
3 wire, 4 wire
After assertion of CS, compares data after arbitrary byte count
and triggers.
Analyzable signals
CH1 to CH4, Logic input, M1 to M2
Byte order
MSB, LSB
Field definition
Field size (4 to 32 bits), Enabled bit range
Analyzable no. of data
300000 bytes max.
List display items
Analysis no., time from trigger position (Time (ms)), Data 1, Data 2
UART Signal Analysis Functions (/F1 and /F3 Options)
*6
Bit rate
115200 bps, 57600 bps, 38400 bps, 19200 bps, 9600 bps,
4800 bps, 2400 bps, 1200 bps, User Define (an arbitrary bit rate
from 1 k to 1 Mbps with resolution of 100 bps)