26
IM AQ6376-02EN
6
Attaching the Connector Adapter
Attach the optional connector adapter before using the instrument.
WARNING
Always turn the power OFF before replacing the connector adapter. The instrument
has a built-in reference light source for wavelength calibration, and infrared light is
always being output from the optical output connector. Never look into the optical
output connector. Infrared light entering the eyes can cause severe injury and loss
of vision.
CAUTION
• As there may be dust adhering to calibration output, be sure to clean it before
attaching the connector adapter.
• Do not exhale or blow compressed air into the monochromator from the optical
input. Doing so may allow dust or other materials to enter the monochromator,
adversely affecting its optical performance. Also, if debris is adhering to the
optical components inside the monochromator when a strong light source is
input, the monochromator may be damaged.
• When attaching or removing the connector adapter, carefully insert it
perpendicularly to the ferrule so as not to damage the ferrule end.
• Moving the connector adapter to the right or left or inserting it forcefully can
damage the ferrule or the connector adapter.
French
AVERTISSEMENT
Toujours éteindre l’avant de remplacer l’adaptateur de connecteur.
Cet instrument dispose d’une source de lumière de référence intégrée pour les
ajustements d’alignement. La lumière infrarouge est toujours émise depuis le
connecteur de sortie optique. Ne regardez jamais directement dans le connecteur
de sortie optique. La lumière infrarouge risquerait de gravement vous blesser ou de
provoquer une perte de vision.