29

IM AQ6376-02EN

Connecting a Communication Interface

Connecting the Mouse

You can use a USB mouse. 

Supported USB Mouse

The instrument can support a USB HID Class Ver. 1.1 compliant mouse (with wheel). 

Connections

The USB mouse is connected to the USB interface on the front panel of the instrument. 

1.

  Confirm that 

MAIN POWER 

switch on the rear panel is OFF. 

2.

  Orient the USB mouse connector in the proper direction so that it is perpendicular 

to the USB port on the front panel, and insert it into the port. 

Note

•  There are two USB ports, but do not connect a mouse to each port at the same time. 
•  In addition to a mouse, you can connect a USB storage device. 

For instructions on using the mouse, see section 3.2 of the 

User’s Manual,  

IM AQ6376-01EN. 

Connecting a Keyboard

You can connect a keyboard for entering file names, comments, and other items. Also, 
the functions and settings of the instrument are assigned to keyboard keys, allowing you 
to manipulate them with a keyboard just as you would by using the instrument’s panel 
keys. 

Supported Keyboards

The instrument supports any 101 English keyboard. 

Connecting

Connect a USB keyboard to one of the USB interfaces on the front or rear panel of the 
instrument.

1.

  Confirm that the 

MAIN POWER

 switch on the rear panel is OFF. 

2.

  Orient the mouse connector so that it matches the orientation of the interface, 

then insert the connector. 

Note

•  There are 2 USB interfaces each on front and rear panels, but do not connect more than 

one keyboard at a time.

•  In addition to a USB keyboard, the USB interfaces can be used to connect USB storage and 

a USB mouse.

For information on operations using the keyboard, see section 3.2 of the 

User’s Manual,

 

IM AQ6376-01EN.

Содержание AQ6376

Страница 1: ...User s Manual AQ6376 Optical Spectrum Analyzer Getting Started Guide IM AQ6376 02EN 6th Edition...

Страница 2: ...w us to serve you best by completing the user registration form accessible from our website https tmi yokogawa com support Contact Us If you want to resolve a technical support issue or need to contac...

Страница 3: ...inted manual This guide explains the handling precautions basic operations and specifications of the AQ6376 Optical Spectrum Analyzer IM AQ6360 92Z1 A document for China Safety Instruction Manual IM 0...

Страница 4: ...2 IM AQ6376 02EN Revisions 1st Edition April 2016 2nd Edition April 2017 3rd Edition October 2017 4th Edition April 2019 5th Edition April 2021 6th Edition December 2021...

Страница 5: ...Rated voltage 250 V N Brazilian standard Rated voltage 250 V T Taiwanese standard Rated voltage 125 V B Indian standard Rated voltage 250 V U IEC Plug Type B Rated voltage 250 V Y No power cord includ...

Страница 6: ...ontains the PDF files shown below The CD also contains Japanese manuals File Name Manual Title Manual No Features Operation Manual pdf AQ6376 Optical Spectrum Analyzer IM AQ6376 01EN User s Manual Com...

Страница 7: ...ly with these requirements The following symbols are used on this instrument Danger Refer to the user s manual This symbol appears on dangerous locations on the instrument which require special instru...

Страница 8: ...strument is a three prong cord Connect the power cord to a properly grounded three prong outlet Do not Impair the Protective Grounding Never cut off the internal or external protective earth wire or d...

Страница 9: ...s d utilisation d une rallonge celle ci doit tre imp rativement reli e la terre Par ailleurs ne pas utiliser ce cordon d alimentation avec d autres instruments Brancher la prise de terre Avant de mett...

Страница 10: ...par cons quent repr sentent un danger Installer et utiliser l instrument aux emplacements appropri s Ne pas installer ni utiliser l instrument l ext rieur ou dans des lieux expos s la pluie ou l eau I...

Страница 11: ...r Products Part 1 Equipment Classification Requirements and User s Guide In addition this instrument complies with 21 CFR 1040 10 and 1040 11 except for deviations pursuant to Laser Notice No 50 dated...

Страница 12: ...ve Regulation valid only in the EEA and UK Battery Regulation in the UK Batteries are included in this product This marking indicates they shall be sorted out and collected as ordained in the EU batte...

Страница 13: ...ser and precautions that can be taken to prevent such occurrences CAUTION Calls attentions to actions or conditions that could cause light injury to the user or damage to the instrument or user s data...

Страница 14: ...re written with inexperienced users in mind experienced users may not need to carry out all the steps Explanation This subsection describes the setup parameters and the limitations on the procedures N...

Страница 15: ...gth Calibration Purging Auto Sweep Setting and Measurement Other Settings Waveform Display Displaying Calculated Waveforms Marker Display Searching Waveform Analysis Waveform Analysis GO NO GO Judgmen...

Страница 16: ...bols and Notation Used in This Manual 11 Flow of Operation 13 1 Front Panel 15 2 Rear Panel 16 3 Panel Keys and Knobs 17 4 LCD Screen 20 5 Installing the Instrument 22 6 Attaching the Connector Adapte...

Страница 17: ...section Used to enter settings pertaining to all measurements sweep measurement conditions data analysis and various functions 4 DATA ENTRY section Used for measurement condition parameter input labe...

Страница 18: ...nnector for synchronous signals for the synchronous measurement function with the Tunable Laser Source 4 ANALOG OUT Analog output 5 MAIN POWER Used to turn the main power ON OFF 6 Power cord connector...

Страница 19: ...unctions related to level axis settings When you press the LEVEL key the soft key menu for setting reference level appears SETUP The SETUP key contains functions related to measurement condition setti...

Страница 20: ...of steps COPY The COPY key is used to output the measurement screen to a file When you press the COPY key the measured waveforms and lists displayed on the screen are output to a file PRESET The PRES...

Страница 21: ...se step for numerical values Each time you press this key the setting toggles between ON and OFF When ON the lamp lights Numeric keypad You can enter numerical values directly into the parameter input...

Страница 22: ...length resolution 9 Displays measurement sensitivity 10 Displays averaging times 11 Displays the number of measurement samples 12 Displays date and time 13 Displays each trace status 14 Only displayed...

Страница 23: ...M AQ6376 02EN 19 Parameter display area 20 Parameter input area 21 OVERVIEW display screen Only displayed when ZOOM function is used 22 Displays sub scale 23 Displayed when SWEEP SPEED is 2x 4 LCD Scr...

Страница 24: ...s greater than or equal to the packaging used upon shipment from the factory Never use inferior packaging materials that are unable to sufficiently absorb vibrations and shocks occurring during transp...

Страница 25: ...le retirer du carton Installation Conditions Desktop Install the instrument so that the following conditions are met Flat Horizontal Location Place the instrument in a stable location that is flat in...

Страница 26: ...attached to the feet at the rear of the instrument to prevent the instrument from sliding One set of rubber stoppers two stoppers are included with the instrument Rubber stoppers A9088ZM Feet at the...

Страница 27: ...ons Do Not Place Anything on Top of the Instrument Never stack instruments or place any other objects on top of the instrument especially those containing water Doing so can lead to malfunction Take P...

Страница 28: ...s to enter the monochromator adversely affecting its optical performance Also if debris is adhering to the optical components inside the monochromator when a strong light source is input the monochrom...

Страница 29: ...mmager la virole ou de l adaptateur de connecteur A connector adapter is required for connecting the optical connector to the AQ6376 On products with the FC SC RFC or RSC option connector adapters com...

Страница 30: ...Latch pin Optical input section Connector adapter Attached Explanation Types of Connector Adapter The connector adapter for internal reference light output AQ9441 comes in the following three types F...

Страница 31: ...nnecting a Keyboard You can connect a keyboard for entering file names comments and other items Also the functions and settings of the instrument are assigned to keyboard keys allowing you to manipula...

Страница 32: ...storage device or turn the power OFF while the USB storage device access indicator is blinking This can damage the data on the device or the device itself French ATTENTION Ne retirez pas le dispositif...

Страница 33: ...outlet that is compatible with the power cord that you will be using and you cannot ground the instrument do not use the instrument French AVERTISSEMENT Assurez vous que la tension d alimentation corr...

Страница 34: ...mum power consumption Approx 100 VA This instrument can use a 100 V or a 200 V power supply The maximum rated voltage differs according to the type of power cord Check that the voltage supplied to the...

Страница 35: ...egins US B PO WER The initialization screen appears and the internal initialization process starts STEP 1 9 through STEP 9 9 are displayed in the lower right part of the screen to indicate the progres...

Страница 36: ...r warm up See section 2 1 of the User s Manual IM AQ6376 01EN for details on the alignment adjustment operation and 2 2 for wavelength calibration When the Power on Operation Does Not Finish Normally...

Страница 37: ...e fichier de configuration du syst me d exploitation ne serait pas sauvegard ce qui entra nerait probablement des dysfonctionnements la prochaine mise sous tension de l instrument Suivez toujours la p...

Страница 38: ...d down the POWER switch for approximately four seconds or longer to force standby mode Note that the operating system configuration file will not be backed up possibly resulting in malfunctions upon s...

Страница 39: ...necting the input light make sure that it does not exceed the AQ6376 s maximum rated level If input light exceeding the maximum rated level is introduced the optical section may be damaged Do not exha...

Страница 40: ...poussi re ou d autres mat riaux pour entrer dans le monochromateur nuire ses performances optiques En outre si des d bris adh re aux composants optiques l int rieur du monochromateur quand une forte s...

Страница 41: ...r cleaner 8 Connect the optical connector on the other end of the optical fiber to the optical connector on the DUT Measuring System AQ6376 Light source Optical fiber Explanation Optical Connectors Ty...

Страница 42: ...nector adapter option required For wavelength reference light source output AQ9441 connector adapter option required Connector types FC or SC Built in calibration light source For alignment and wavele...

Страница 43: ...display calculation between traces display normalized display roll averaging sweep average display 2 to 100 times curve fit display peak curve fit display marker curve fit display Other Trace copy tr...

Страница 44: ...ore 4 When using 9 5 125 m single mode fiber SSMA type in JIS C6835 PC polishing mode field diameter 9 5 m NA 0 104 to 0 107 5 After wavelength calibration with built in reference light source samplin...

Страница 45: ...ternal Dimensions Unit mm approx inch 426 12 3 12 3 20 8 221 14 8 458 8 32 0 48 16 77 0 48 0 82 8 70 0 58 18 06 1 26 REAR VIEW If not specified the tolerance is 3 However in cases of less than 10 mm t...

Отзывы: