Chapter 6
Maintenance, Inspection and Storage
6.3
Troubleshooting
6-5
6
6.3
Troubleshooting
Problem
Probable Cause
Corrective Action
Reference Page
Using power supply outside
the ratings.
Use a proper supply.
3-2 7-2
Connector wiring is not
correct.
Check that the wiring is
correct.
ⅷ
The power does not
turn ON.
Fuses are not set.
Set fuses.
ⅹ
Shutter is closed.
Push [OUT] switch.
2-2
Fiber is disconnected.
Exchange the fiber.
4-2
The output level is not
correct.
Lamp is damaged.
Replace the lamp.
2-5
Dust adhere to the fiber.
Clean the fiber.
4-2 6-4
Lamp is damaged.
Replace the lamp.
2-5
The output level is
instability.
Environmental conditions are
no good.
Check the environmental
conditions.
7-2
Parameter settings are not
matched.
Match the parameters.
5-3
Cannot control the
instrument via the GP-
IB interface.
Make a mistake in GP-IB
address
Check GP-IB address.
2-3 5-3
6.4
Care during Storage
If the unit is stored for a long time, sufficient care should be taken of the following:
①
Ensure that the unit has been wiped clean of fingerprints, dust, etc.
②
Perform instructions given in Section 6.2 Operational Inspection to check that the unit
operates properly.
③
Do not store the unit in the following locations:
✓
where it would be exposed to direct sunlight or excessive dust
✓ where there is
high humidity that could result in the formation of water droplets or
exposure to them
✓
where it would be exposed to activated gas or the possibility of oxidization
✓
where temperature and humidity conditions are as shown below:
・
Temperature >50
℃
, <-10
℃
・
Humidity >80%
For extended storage, it is desirable that the unit be stored within the range of the following
environmental conditions while, at the same time, the above conditions are met.
✓
Temperature 5 - 30°C
✓ Humidity
40 - 70%
✓ Daily fluctuations of temperature/humidity are small.
z
When the unit is used again after storage, perform instructions given in
Section 6.2 "Operational Inspection" to check that it operates properly.
(
☛
Section 6.2 Operational Inspection)
z
To resume usage, xyz-axes adjustment procedures must be performed.
(
☛
Sections 4.3 Xyz-axes Adjustment)
Caution
Содержание AQ4305
Страница 1: ......
Страница 2: ......
Страница 4: ...ii ...
Страница 14: ...CONTENTS xii ...
Страница 20: ...1 4 Warranty Chapter 1 Introduction 1 6 ...
Страница 30: ...3 2 Interface Connection Procedures Chapter 3 Before You Start 3 4 ...
Страница 36: ...4 3 Xyz axes Adjustment Chapter 4 Usage Preparations 4 6 Fig 4 5 The optical output level vs Z axis deviation ...
Страница 40: ...5 2 GP IB interface Chapter 5 Operating Procedures 5 4 ...
Страница 50: ...77 2 External Dimensions Chapter 7 Specifications 7 4 ...