YOKOGAWA Admag AXW Скачать руководство пользователя страница 17

 

 

<3.  INSTALLATION>

3-8

IM 01E25D11-01EN

Table 3.3.1  Tightening Torque Values for Metal Piping (N-m)

Unit: N-m

Lining type

PTFE / Natural hard rubber /

 

Natural soft rubber

Gasket type

Soft rubber gasket, or the equivalent in hardness (supplied by customer)

Process

connection

Size

mm (inch)

ASME

B16.5 (500, 600)

B16.47 (700-1000) 

Class 150

EN1092-1 PN10

JIS B2220 10K

EN 1092-1 PN16

AWWA C207

Class D

AS2129 Table D, E

AS4087 PN16

500 (20)

99 to 166

85 to 140

143 to 238

151 to 233

600 (24)

144 to 240

108 to 175

218 to 363

239 to 331

700 (28)

168 to 281

176 to 266

180 to 300

176 to 293

240 to 333

800 (32)

259 to 432

212 to 302

236 to 393

246 to 410

400 to 612

900 (36)

274 to 457

206 to 338

224 to 373

258 to 430

399 to 604

1000 (40)

274 to 457

277 to 422

307 to 512

286 to 477

422 to 587

Unit: N-m

Lining type

Polyurethane rubber

Gasket type

Soft rubber gasket, or the equivalent in hardness (supplied by customer)

Process

connection

Size

mm (inch)

ASME

B16.5 (500, 600)

B16.47 (700-1000)

Class 150

EN1092-1 PN10

JIS B2220 10K

EN 1092-1 

PN16

AWWA C207

Class D

AS2129 Table 

D, E

AS4087 PN16

JIS F12

(JIS 75M)

500 (20)

71 to 118

60 to 99

74 to 123

76 to 123

131 to 218

600 (24)

97 to 162

82 to 126

101 to 168

118 to 171

124 to 206

700 (28)

125 to 208

135 to 202

146 to 244

120 to 200

167 to 242

202 to 337

800 (32)

176 to 294

160 to 229

190 to 316

169 to 281

218 to 348

197 to 328

900 (36)

170 to 283

151 to 251

178 to 296

161 to 268

200 to 317

221 to 369

1000 (40)

175 to 291

206 to 314

220 to 367

165 to 275

253 to 370

218 to 363

Unit: N-m

Lining type

Natural hard rubber

Polyurethane 

rubber

Gasket type

Soft rubber gasket, or the equivalent in hardness (supplied by customer)

Process

connection

Size

mm (inch)

EN1092-1 PN6

EN1092-1 PN10

AWWA C207

Class D

AS2129 Table D, E

AS4087 PN16

JIS F12

(JIS 75M)

1100 (44)

285 to 476

1200 (48)

180 to 300

327 to 546

298 to 497

400 to 582

281 to 468

1350 (54)

397 to 662

1400 (56)

233 to 388

379 to 632

1500 (60)

421 to 702

1600 (64)

238 to 397

532 to 886

532 to 887

1800 (72)

291 to 485

552 to 919

468 to 779

Содержание Admag AXW

Страница 1: ...User s Manual AXW Magnetic Flowmeter Remote Flowtube IM 01E25D11 01EN IM 01E25D11 01EN 2nd Edition...

Страница 2: ...ns 3 1 3 2 Handling Precautions 3 3 3 2 1 General Precautions 3 3 3 2 2 Flowmeter Piping 3 4 3 3 Mounting Procedures 3 4 4 WIRING 4 1 4 1 Wiring the Remote Flowtube 4 1 4 1 1 Wiring Precautions 4 1 4...

Страница 3: ...ii IM 01E25D11 01EN n OPTIONAL SPECIFICATIONS 6 5 n OPTIONAL SPECIFICATIONS continued 6 6 n DIMENSIONAL DRAWINGS 6 7 n SIZING DATA 6 19 Revision Information i...

Страница 4: ...t changes that are not considered to affect function or performance If the customer or any third party is harmed by the use of this product Yokogawa assumes no responsibility for any such harm owing t...

Страница 5: ...2 Wiring WARNING The wiring of the magnetic flowmeter must be performed by expert engineer or skilled personnel No operator shall be permitted to perform procedures relating to wiring When connecting...

Страница 6: ...ould appreciate the inclusion of drawings or additional information The results of our examination will determine whether the meter will be repaired free of charge or on an at cost basis The guarantee...

Страница 7: ...Blank Page...

Страница 8: ...A place where it will not be exposed to rain or water A place subject to minimal vibrations or shocks Temperature and humidity levels should be as follows Temperature 10 to 70 C Humidity 5 to 80 RH n...

Страница 9: ...Blank Page...

Страница 10: ...ng two or more magnetic flowmeters on a single pipe provide a run of at least 10D between them F0301 ai D Flowtube Size Gate valve fully open 5D or more 2D or more Reducer pipe Expander pipe 0 is allo...

Страница 11: ...r to the pipeline set a expansive pipe in the downstream side 9 Mounting Positions Pipes must be fully filled with liquids IMPORTANT It is essential that pipes remain fully filled at all times otherwi...

Страница 12: ...ed to the installing location transport it to the site in its original packing CAUTION When lifting the flowtube use the lifting rings eye bolts or eye plates as in Figure 3 2 1 Never lift it using a...

Страница 13: ...substances passing through the flowtube during measurement 3 3 Mounting Procedures CAUTION When attaching optional grounding rings to the flowtube with lining material natural hard rubber or fluoroca...

Страница 14: ...of gasket should be a kind of soft rubber or its equivalent in hardness 1 Standard no grounding rings Gaskets A supplied by customer Customer pipe Customer pipe A A F0310 ai Figure 3 3 2 Installation...

Страница 15: ...of the flowtube s flange and fix the wire by the nut This procedure item 1 to 4 must be done for the both sides of the flowtube 5 Install the flowtube into the customer s pipe with the gaskets A F0313...

Страница 16: ...Procedure with Grounding Rings GR1 for sizes 1100 mm 44 in and above Mounting Procedure with optional Grounding Rings GR2 The grounding rings are installed and locked onto the flowtube with gaskets B...

Страница 17: ...ess connection Size mm inch ASME B16 5 500 600 B16 47 700 1000 Class 150 EN1092 1 PN10 JIS B2220 10K EN 1092 1 PN16 AWWA C207 Class D AS2129 Table D E AS4087 PN16 JIS F12 JIS 75M 500 20 71 to 118 60 t...

Страница 18: ...mm inch ASME B16 5 500 600 B16 47 700 1000 Class 150 EN1092 1 PN10 JIS B2220 10K EN 1092 1 PN16 AWWA C207 Class D AS2129 Table D E AS4087 PN16 JIS F12 JIS 75M 500 20 626 to 1044 527 to 876 652 to 108...

Страница 19: ...Blank Page...

Страница 20: ...gth All the cable ends must be provided with round crimp on terminals and be securely wired The signal cables must be routed in separate steel conduit tubes 16 JIS C 8305 or flexible conduit tubes 15...

Страница 21: ...rter side On the flowtube side Figure 4 1 2 Treatment of Dedicated Signal Cables CAUTION As crimp terminals A B SA SB and C have their own electrical potentials securely insulate them so as not to com...

Страница 22: ...ing at this terminal box is not necessary and do not open it IMPORTANT The wiring port is sealed with a cap not water proof Do not remove the cap from the wiring port until wiring a cable If waterproo...

Страница 23: ...regularly Drain valve F0406 ai Figure 4 1 6 Conduit Wiring 4 1 4 Wiring Connections WARNING Before wiring be sure that the power supply for AXFA11 converter has been turned off to prevent an electric...

Страница 24: ...grounding can have an adverse affect on the flow measurement Ensure that the instrument is properly grounded The electromotive force of the magnetic flowmeter is minute and it is easy to be affected...

Страница 25: ...Blank Page...

Страница 26: ...ot be changed direction of electrical connection after delivery Submersible Type Optional code DHC for district heating and cooling or condensation proof 1 The following tools are required to change t...

Страница 27: ...an abnormal condition Consult Yokogawa s sales or service offices 2 Insulation Resistance Check Check the insulation resistances in the terminal box in accordance with the tables below If any of them...

Страница 28: ...flowtube is filled with fluid Is the flowtube full of fluid Remove the tube and inspect the inside of the pipe Is foreign matter adhering to the walls Does the arrow on the tube match the flow directi...

Страница 29: ...imits Whenever possible reduce the flow velocity to being in the vicinity of the lower conductivity limit Remove dirt and scale of and near electrodes Use a soft brush or cloth to perform scale remova...

Страница 30: ...ed by turbulence in the flow Relocate the valve to the downstream side If the valve must be located on the upstream side ensure that the minimum distance to the flowmeter is 10 times the flow tube dia...

Страница 31: ...Blank Page...

Страница 32: ...een Submersible type Pipe Housing Flange Terminal Box Non tar epoxy coating black Wiring Electrical connection ANSI 1 2 NPT female ISO M20 x 1 5 female JIS G1 2 PF1 2 female Direction of Electrical Co...

Страница 33: ...rature 10 to 60 C 14 to 140 F Note Minimum temperature should also be limited accord ing to minimum fluid temperature of flow tube s speci fication Refer to description of Fluid Temperature and Pressu...

Страница 34: ...p to 0 6 MPa CE1 EN1092 1 PN10 Flange Carbon Steel fluid pressure up to 1 MPa 500 mm 20 in to 1000 mm 40 in for lining code F H D U 1200 mm 48 in to 1800 mm 72 in for lining code H CE2 EN1092 1 PN16 F...

Страница 35: ...G1 2 female 2 ANSI 1 2 NPT female Not available for Submersible type 4 ISO M20 x 1 5 female Not available for Submersible type Indicator N None Calibration B Standard Optional Specifications Optional...

Страница 36: ...zer display pulse weight a Available density Numerics Specify the numeric within the value of 0 0001 to 32000 And it can be up to five digits to a maximum of 32000 ignoring the decimal point A fractio...

Страница 37: ...ard List are issued L3 Level 4 The Declaration and the Yokogawa Measuring Instruments Control System are issued L4 Five point Calibra tion in User specified Span A flow test near 0 25 50 75 and 100 of...

Страница 38: ...749 3 29 50 863 6 34 00 977 9 38 50 1086 42 75 1200 47 25 Inner Diameter d Lining H F D 477 18 78 578 22 76 675 26 57 777 30 59 878 34 57 976 38 43 Lining U 469 18 46 570 22 44 667 26 26 769 30 28 870...

Страница 39: ...ble 3 EN PN10 Flange Type Unit mm approx in Process Connection Code CE1 Flange Type EN1092 1 PN10 Size Code 500 600 700 800 900 10L Size 500 20 600 24 700 28 800 32 900 36 1000 40 Lining Code H F U D...

Страница 40: ...ion code DHC is selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 5 JIS 10K Flange Type Unit mm approx in Process Connection Code CJ1 Fl...

Страница 41: ...s selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 7 AS2129 Table E Flange Type Unit mm approx in Process Connection Code CS2 Flange Ty...

Страница 42: ...submersible type or option code DHC is selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 9 JIS F12 JIS 75M Flange Type Unit mm approx in...

Страница 43: ...24 700 28 800 32 900 36 1000 40 Lining Code H F U D Ring Outer Diameter D 590 23 23 690 27 17 799 31 46 906 35 67 1006 39 61 1116 43 94 Ring Inner Diameter d 485 19 09 586 23 07 683 26 89 787 30 98 8...

Страница 44: ...ion Code CB1 Flange Type AWWA C207 Class D Size Code 12L Size 1200 48 Lining Code H Lay Length L 1440 56 69 Flange Outer Diameter D 1511 3 59 50 Flange Bolt Circle Diameter C1 1422 4 56 00 Ring Bolt C...

Страница 45: ...rproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 3 EN PN10 Flange Type fluid pressure up to 0 6 MPa Unit mm approx in Process Connection Code CEK Fl...

Страница 46: ...type or option code DHC is selected waterproof glands and a 30 meter length cable are attached Add 9 5 kg 20 9 lb to the weight in the table 5 AS2129 Table D AS2129 Table E AS4087 PN16 Flange Type Uni...

Страница 47: ...Diameter d Lining U 1071 42 17 1172 46 14 1322 52 05 1469 57 83 1565 61 61 1759 69 25 Bolt Hole Pitch Half Angle 7 5 6 4 6 4 5 6 5 4 1 Bolt Hole Diameter h 33 1 30 33 1 30 39 1 54 39 1 54 39 1 54 39...

Страница 48: ...516 59 69 1716 67 56 1925 75 79 1335 52 56 1540 60 63 1766 69 53 1966 77 40 Ring Inner Diameter d 1195 47 05 1201 47 28 1404 55 28 1592 62 68 1790 70 47 1195 47 05 1396 54 96 1588 62 52 1782 70 16 Bol...

Страница 49: ...Size Code 11L 12L 13L 15L 16L 18L Size 1100 44 1200 48 1350 54 1500 60 1600 64 1800 72 Lining Code U Thickness 25 0 98 Approx Weight Unit kg lb 59 3 131 65 7 145 75 8 167 88 3 195 92 4 204 114 2 252 O...

Страница 50: ...95 0 08 2 3 0 09 2 7 0 11 3 15 0 12 3 6 0 14 4 05 0 16 4 45 0 18 4 85 0 19 500 19 69 630 24 80 800 31 50 1000 39 37 1250 49 21 1600 62 99 2000 78 74 2500 98 43 630 24 80 800 31 50 1000 39 37 1250 49...

Страница 51: ...Blank Page...

Страница 52: ...le AXW Magnetic Flowmeter Remote Flowtube Manual No IM 01E25D11 01EN Edition Data Page Revised Item 1st Mar 2012 New publication 2nd Sep 2013 Addition of sizes from 1100 mm 44 in to 1800 mm 72 in Addi...

Отзывы: