48
IM 702915-01EN
Overload
POWER
OVERLOAD
JAW UNLOCKED
RANGE
30A
0.1V/A
5A
1V/A
0.5A
10V/A
AUTO ZERO
DEMAG
Lights in green
Blinks rapidly in red
turns off
One LED lights in green
Lights in red or turns off
A current exceeding the following level specified for each current range is
detected.
30 A range:
32.5 ±2.5 Arms
5 A range:
5.25 ±0.25 Arms
0.5 A range:
0.525 ±0.025 Arms
(DC and 45 to 66 Hz sine waves)
If using the 30 A range, remove the sensor from the conductor to be
measured. For the 0.5 A or 5 A range, switch to the larger current range.
Overload cannot be detected for currents momentarily exceeding the
specified level or for high-frequency currents exceeding the specified
level. Also, immediately after switching the current range, overload may
not be detected correctly. The OVERLOAD LED may blink due to the
influence of an external magnetic field when the jaws are open, but this is
not a malfunction.
6. Troubleshooting