49
IM 702915-01EN
Demagnetizing / Automatic Zero-Adjustment Error
POWER
OVERLOAD
JAW UNLOCKED
RANGE
30A
0.1V/A
5A
1V/A
0.5A
10V/A
AUTO ZERO
DEMAG
Lights in green
Blink 3 times in red
Blinks 3 times in orange
One LED lights in green
Demagnetizing and automatic zero-adjustment functions cannot be
executed. Alternatively, demagnetizing or automatic zero-adjustment has
not finished properly. The instrument is in one of the following status.
• The upper jaw not locked. (JAW UNLOCKED lights up.)
• An overload detected regardless of Jaw locked state. (OVERLOAD
blinks rapidly.)
• A measured current exceeding the following values detected regardless
of Jaw locked state.
0.50 ± 0.25 Arms (DC, 45 to 66 Hz sine waves)
Take any of the following actions, depending on the status after the LEDs
blink 3 times. After that, demagnetizing and automatic zero-adjustment
will be performed again.
6. Troubleshooting