• The measured value may be affected by the conductor position within
the sensor aperture. Position the conductor at the center of the sensor
window, and make the straight portion as long as possible. Avoid
positioning the conductor so that it is wound locally around the sensor
head.
• To make accurate measurement, hold down the opening lever until
the JAW UNLOCKED indicator disappears to ensure that the sensor
head is securely closed and locked.
• Conductor clamping and cable positioning may cause load
fluctuations, affecting the observed waveform.
• To prevent offset voltage variation from occurring, make sure that
force is not applied to the jaw or opening lever during measurement.
Ending Measurement
Notes on Ending Measurement
To remove the instrument’s power cable from the power supply, pull the
power cable shell.
CAUTION
• To prevent breaking the cable, do not pull the cord when
removing the output terminal from the waveform measuring
instrument. Be sure remove by holding the terminator and
pulling the lock release lever toward you.
• To prevent breaking the instrument’s power cable, probe
power supply terminal of the waveform measuring instrument,
and the power supply receptacle of the 701934 power supply
unit, observe the following when removing the power cable.
• Hold the power plug shell and remove.
• Do not pull the cable or rotate the power plug. (Pulling the
power plug shell releases the lock allowing you to remove it
from the power supply receptacle.)