Voltage, Current, and Resistance Sourcing
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Output Setting Display
• The output setting appears flush right at the resolution of each range.
• The ratio of the output setting to the range is passed on to the output setting even when you change
the range.
Example: If the output setting is 1.5000 V (15%) at the 10 V range and you change the voltage
range to 30 V, the output setting will be 4.500 V.
10 V Range
30 V Range
• When you change the range and the specified output setting is outside the selectable range of the
new range, the output setting and polarity displays blink. If you turn any of the output setting dials
in this condition but the output setting still remains outside the selectable range of the new range,
the output setting is set to the maximum or minimum setting of the new range, and the blinking will
change to solid.
<<Command Mnemonic>>
:OUTPut:POLarity {NORMal|INVerted}
:SOURce:FUNCtion {VOLTage|CURRent|RESistance}
:SOURce:LEVel {<
Voltage
>|<
Current
|<
Resistance
>|MINimum|MAXimum}
:SOURce:SLEVel {<
Voltage
>|<
Current
|<
Resistance
>|MINimum|MAXimum}
5.2 Specifying the Output Setting and Polarity