YL9112 Isocratic Pump
5-4. Replacement of check valve cartridge
If check valve is not well operated due to contamination, pressure change is severe during operation and
pump does not operate properly. Many problems of check valve are caused by small impurity that
interferes with operation of check valve. Therefore, if impurity is prevented from entering inside of pump
head using mobile phase filter, malfunction of check valve is almost not caused. The cleaning of the check
valve cartridges using sonication can solve this problem.
1) Separate tubing connected to pump head.
2) Loosen inlet and outlet check valve housing of pump head using spanner.
3) Wash check valve cartridge in separated check valve housing for about 30 minutes using
ultrasonic cleaner with 10% nitric acid solution.
4) Using pure water, rinse check valve cartridge to remove the nitric acid used for cleaning.
5) Assemble loosened check valve in the reverse order.
5-5. Replacement of low pressure seal
Wearing of the low pressure seal is caused when pump has been used for long time without using
washing port. In order to prevent wear of low pressure seal, it is desired to use washing port, and it is
more desirable in case of using buffer solution. Leakage of liquid due to wearing of the low pressure seal
is caused between pump head and body.
1) Separate pump head with reference to 5.2.
2) Separate washer, and pull out low pressure seal assembly from pump headbody with seal
insertion/removal tool.
3) Replace low pressure seal attached to low pressure seal assembly with new one.
4) Assemble pump head in reverse order by referring to 5.2.
5-6. Cleaning of flow path within pump
In order to prevent occurrence of problem in instrument, remove impurity accumulated in instrument,
and it is better to clean flow path when it is not used for long time. Clean inside of flow path in the following
26
Содержание YL9100 HPLC
Страница 1: ......
Страница 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Страница 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Страница 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Страница 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Страница 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Страница 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Страница 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Страница 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Страница 188: ......