YL9112 Isocratic Pump
YL9112 isocratic pump provides output signal(12V 500mA) to operate external solenoid valve as like
[Fig 12].
Y
L9112
[Fig. 12] Diagram of solenoid output
[Remote operation]
START-IN
: Operate instrument, and start running of gradient program.
If you connect it with autosampler or external valve, automatic running is available.
START-OUT : If the signal input on the START-IN terminal, the signal pulse output through this port. It can
be used for synchronization of remote start with the other instrument.
MARK-IN
: To control event program or operate additional operation.
MARK-OUT
: To control time event program output.
READY-IN : To change error state and stop operation if there is a input.
READY-OUT : When YL9112 isocratic pump is not ready state because of running status, output
error signal is indicated if there is a leak.
3-5. Connection of communication cable
YL9112 isocratic pump provides TCP/IP internet protocol as a standard. The IP address of YL9100
pump series is 10.10.10.10, if DIP SW settings on the rear side are On position. If you change the IP
address using control software, the DIP SW has to be set OFF.
13
Содержание YL9100 HPLC
Страница 1: ......
Страница 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Страница 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Страница 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Страница 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Страница 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Страница 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Страница 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Страница 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Страница 188: ......