![Yamato WG 1000 Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/yamato/wg-1000/wg-1000_instruction-manual_905422035.webp)
32
Maintenance Method
Daily Inspection and Maintenance
Timing of maintenance and inspection
(Perform daily inspection for stable use of product.)
Maintenance/check items
Reference for timing
Remarks
Replacement of
pre-treatment cartridge
Approx. 6 months
Exchange message appears
on the indication area.
Throughput: About 5000 liters of city water
in Tokyo.
If the quality of raw water is not good,
perform this maintenance as soon as
possible.
Replacement of
membrane filter
Approx. 3 months
Exchange message appears
on the indication area.
Throughput: approx. 500 liters for pure
water.
Need to exchange if the amount of collection
is less than the above value.
Replacement of ion exchange
resin cartridge
Exchange message appears
on the indication area.
(Displayed when the
conductivity of pure water is
more than 1 x 10
-4
S/m.)
Throughput: About 700 liters of the raw
water of 200 x 10
-4
S/m. (1400 liter for
WG1000 type)
Cleaning of distiller
3 months
Cleaning of water supply
hose filer
6 months
If the quality of raw water is not good, start
the operation as soon as possible.
Replacement of hose
2 years
Connections shall be checked monthly.
Drainage of distilled water tank
3 months
If not used for a long time, drain water from
the tank.
Replacement of pre-treatment cartridge
See Page 11 "21. Secure the pre-treatment cartridge securely" for replacement procedure.
When the cartridge is put to use without replacement, life span of ion exchange resin cartridge
becomes short.
Please dispose used cartridge as nonconbustibles. When it is sent back to our company, fill in the
designated invoice attached to a new cartridge, and send it with the used cartridge to us.
Replacement of ion exchange resin cartridge
When a cartridge is stored for a long time, deterioration of water quality and fall of processing
capacity are found; therefore prepare a spare cartridge in a planned manner for replacement
timing. Standard for storage is about 4 months.
In replacement, see Page 10 "20. Install the ion exchange resin cartridge (CPC-S) securely".
When a cartridge is used without replacement, much scale is deposited on boiler and heater,
which causes decrease of distilled water sampling and damage to heater.
Please dispose used cartridge as nonconbustibles. When it is sent back to our company, fill in the
designated invoice attached to a new cartridge, and send it with the used cartridge to us.
We promote the reasonable disposal, collection, and recycling of cartridge for environmental
protection.
※
After changing the pretreatment cartridge, the membrane filter, or the ion-exchange resin cartridge,
press and hold the reset switch at backside of the door for two seconds till a beep is heard. Then the
warning indication is reset and the change record is registered in the consumables change history.
In case that two or more items are changed at the same time, they are registered in the consumables
change history in order of occurrence. However, be sure that the ion-exchange resin cartridge is
normally registered at the last. (Warning alarm for the ion-exchange resin cartridge is automatically
recovered.)