Before Using this unit
Requirements for Installation
21. Secure the pre-treatment cartridge securely
•
Connect the hose in the body securely following the procedure shown below.
•
Insecure connection may cause disconnection of connection hose, resulting in accident by
water leakage.
1. Make sure that the earth leakage breaker of this unit is "OFF" and that the tap is tightened.
2. Take the pre-treatment cartridge
①
out of attachments to the body.
3. Inlet and outlet of the pre-treatment cartridge
①
are provided with a cap, so remove it.
4. When the front door of this unit is opened, connection hose marked IN and OUT is found in
the coupler; there make connection matching them with IN and OUT on pre-treatment
cartridge
①
.
5. In connecting, mate the coupler and port of cartridge while sliding the blue portion
③
of
coupler toward the hose, push in, then release the blue portion
③
.
6. When connection is finished, place the pre-treatment cartridge at the position shown on the
right (near side on the left of distilled water tank) as paying attention to the bend of hose.
Fig.5
Fig.6
22. After installation
•
This unit may topple over due to unexpected earthquake or shock causing injury. Take an
appropriate measure against toppling for safety.
(For WG1000, earthquake-resistant fittings are optionally available.)
②
①