8. TROUBLESHOOTING
Troubleshooting Guide
Troubleshooting table
Symptom
Check Points
Display does not come up when
main power switch is turned on.
・
Is the power cable connected?
・
Is a power outage in progress?
・
Is power source voltage adequate?
Air does not purge.
Safety valve is activated.
・
Is the exhaust hose kinked or clogged?
・
Is the line clogged with debris?
Water does not drain.
Water does not drain in drying
process.
・
Is drain filter clogged?
Sterilization temperature does rise.
Pressure does not build.
・
Is the lid closed securely?
・
Is seal or flange damaged?
Pressure increases when solenoid
valve is open.
・
Is the chamber exhaust port clogged?
Temperature fluctuates during
operation.
・
Are there considerable changes in external temperature?
Steam discharges rapidly.
・
Is there water in the drain bottle?
・
Is exhaust hose positioned incorrectly or damaged?
Water leaks.
・
Is drain valve closed all the way?
・
Is the drain bottle overfilled?
Operation does not start from
stand-by.
・
Is drain bottle in position?
Loud noise during air purge.
・
Is the silencer installed correctly?
◆
For further technical assistance, contact Yamato Scientific at (800) 292-6286