2. PRE-OPERATION PROCEDURES
Installation Procedures
17. Install drain bottle.
For the safety of equipment and operators, SM series units feature a safety switch, which does not
allow the the sterilizer to run without the drain bottle installed. Always be sure the drain bottle is
installed before each operation. Do not attempt to operate unit without the bottle installed and do not
attempt to modify or remove the drain bottle safety switch.
①
Open the front panel door. The display will
flash
, indicating drain bottle is not
installed.
②
Insert the end of the drain hose, with silencer
attached, into the drain bottle containing
1500cc of water.
③
While lifting the bottle into place, insert the
drain hose the rest of the way into the bottle.
④
Once the bottle is in place, the safety switch is
depressed, allowing the sterilizer to be run.